• J. Nucl. Med. · Jun 2000

    PET with a dual-head coincidence camera: spatial resolution, scatter fraction, and sensitivity.

    • W D Kunze, M Baehre, and E Richter.
    • Clinic of Radiotherapy and Nuclear Medicine, Medical University of Luebeck, Germany.
    • J. Nucl. Med. 2000 Jun 1; 41 (6): 1067-74.

    UnlabelledScintillation cameras with options for detecting positron annihilation quanta in the coincidence acquisition mode may be the most innovative diagnostic devices introduced in nuclear medicine during the last few years. Besides conventional low-energy imaging in the collimated single-photon mode, these options offer a relatively inexpensive opportunity to perform uncollimated PET by switching into the coincidence acquisition mode. Instead of collimators, scatter frames (with 2 optional configurations: axial or open scatter frame) can be mounted to reduce the amount of quanta reaching the detectors from parts of the patient's body outside the field of view. This study investigates the coincidence imaging properties of the scintillation camera by measuring spatial resolution, scatter fraction, sensitivity, and count-rate response for 18F.MethodsA needle in air and a plastic tube in water, each filled with 18F, were oriented axially and transversally to measure the transverse and axial spatial resolutions, respectively. Using either the axial or the open scatter frame, a standard cylinder filled homogeneously with activity was studied over several half-life periods to deduce the respective scatter and random fractions of the system by means of a sinogram technique. The activity of the cylinder was kept low to determine the sensitivity to coincidence events for both scatter frames.ResultsDepending on the distance between the line source and the axis of rotation and on the choice of the axial acceptance angle used to reconstruct the coincidence events (single-slice rebinning algorithm), the axial resolution was found to be between 6 and 10 mm (full width at half maximum) with the axial scatter frame mounted. The transversal resolution was 6-6.5 mm on the axis of rotation, independent of the scatter frame used. The scatter fraction amounted to roughly 25% for the axial and 38% for the open scatter frames. The sensitivity when measuring true coincidence pairs ran to nearly 650 Hz/kBq/mL, when acquisition was performed with the axial scatter frame using a 30%-wide photopeak energy window. When acquiring with the open scatter frame, the sensitivity increased to nearly 3000 Hz/kBq/mL. Using the axial scatter frame, the homogeneously filled cylinder could be scanned with a maximum true coincidence rate of 2000 Hz for an activity of 55-60 MBq. Although this maximum true coincidence counting rate did not change significantly when the acquisition was performed with the open scatter frame, the respective activity in the standard cylinder was decreased to 10-15 MBq.ConclusionThe spatial resolution of the scintillation camera is sufficient for high-resolution coincidence imaging. Compared with a dedicated PET scanner, the scatter fraction is relatively high and should therefore be corrected adequately. The relatively low sensitivity and the rather low maximum true coincidence counting rate are considerably inferior compared with a conventional PET scanner. However, these drawbacks can be partially compensated for, facilitating its clinical use.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…