• Investigative radiology · Jan 2015

    Impact of measurement parameters on apparent diffusion coefficient quantification in diffusion-weighted-magnetic resonance imaging.

    • Holger Schmidt, Sergios Gatidis, Nina F Schwenzer, and Petros Martirosian.
    • From the *Department of Radiology, Diagnostic and Interventional Radiology, and †Department of Radiology, Diagnostic and Interventional Radiology, Section on Experimental Radiology, Eberhard Karls University, Tübingen, Germany.
    • Invest Radiol. 2015 Jan 1; 50 (1): 46-56.

    ObjectiveThe scope of this work was to systematically evaluate the reproducibility of diffusion-weighted imaging and the impact of b values used for apparent diffusion coefficient (ADC) calculation as well as the echo time (TE) on the resulting ADC in phantom studies. We attempted to find a minimum upper b value needed for reliable ADC measurements. In addition, we were able to investigate these impacts not only for different diffusivities but also for different T2 relaxation times. The influence of different b values on ADC calculations for different organs was also assessed in a volunteer study.Materials And MethodsDiffusion-weighted imaging of a phantom consisting of 16 compartments with combinations of 4 different diffusivities and 4 different T2 relaxation times was conducted 5 times using 11 b values (0-1000 s/mm) and 5 different TEs. Apparent diffusion coefficient was calculated from the 16 compartment regions of interest using 42 different combinations of b values. Reproducibility of ADC was assessed from the coefficient of variation of the 5 measurements. The ADC stability was determined from a voxel-based coefficient of variation (CVsta) and the signal-to-noise ratio (SNR) to find the minimum upper b values for a reliable ADC quantification. The influence of TE on ADC quantification was assessed for 9 different b value combinations. The influence of 9 different b value combinations on ADC was evaluated by a region of interest analysis of 7 organs in 12 volunteers.ResultsThe found coefficient of variation was between 10.2% and 1.4%, decreasing with increasing upper b value and increasing diffusivities. Accordingly, CVsta and SNR showed the same trend. Using an upper b value of 600 s/mm gives already reliable ADC results showing a maximum CVsta of 7.5%, whereas an upper b value of 1000 s/mm revealed a maximum CVsta of 5.5%. Values of ADC reduced with increasing upper b value in phantom as well as in human data. Apparent diffusion coefficient also reduced with increasing TE and tended to increase for increasing T2 relaxation times and increasing diffusion restriction.ConclusionsApparent diffusion coefficient can be measured with high reproducibility but strongly depends on b values used and TE, which should be kept constant in each examination protocol. Whereas upper b values as low as 400 s/mm can be used for examinations of tissues with low diffusivities, very high b values (>1000 s/mm) are needed to reach an optimal SNR for high diffusive tissues. An upper b value of 600 s/mm is a good compromise regarding ADC stability, SNR, and measurement time for all tissue types.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.