J Magn Reson Imaging · Jul 2003
Comparative StudyA single breath-hold multiecho T2* cardiovascular magnetic resonance technique for diagnosis of myocardial iron overload.
- Mark Westwood, Lisa J Anderson, David N Firmin, Peter D Gatehouse, Clare C Charrier, Beatrix Wonke, and Dudley J Pennell.
- Cardiovascular MR Unit, Royal Brompton Hospital and Imperial College, London, UK. m.westwood@rbh.nthames.nhs.uk
- J Magn Reson Imaging. 2003 Jul 1; 18 (1): 33-9.
PurposeTo assess tissue iron concentrations by the use of a gradient echo T2* multiecho technique.Materials And MethodsWe compared the results of measurements of heart T2* from 32 patients using the established multiple breath-hold variable TR technique with a new multiecho sequence that acquires all images within a single breath-hold with constant TR.ResultsThere was good agreement of myocardial T2* values between both methods in the abnormal range of T2* < 20 msec (mean difference 0.2 msec, 95% CI -1.3 to 0.9 msec, r = 0.97, P < 0.0001). The coefficient of variability between the methods was 3.5%. The interstudy reproducibility using the multiecho sequence had a variability coefficient of 2.3% in the abnormal T2* range and 5.8% over all T2* values. There was good agreement between the techniques for the liver T2* values.ConclusionsThe use of the single breath-hold, multiecho acquisition allowed reliable quantification of myocardial T2*. The good reproducibility, speed, and T1 independence of this technique allows greater accuracy, faster patient throughput, and, therefore, reduced costs (which is important in developing countries where thalassemia is most prevalent).Copyright 2003 Wiley-Liss, Inc.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.