    Performance Evaluation of Nasal Prong Interface for CPAP Delivery on a Critical Care Ventilator: A Bench Experiment.

    • Natalie Napolitano, Tracey Roberts, Amanda J Nickel, Joseph McDonough, Haorui Sun, Rui Feng, Erik A Jensen, Kevin Dysart, and Richard Lin.
    • Department of Respiratory Therapy, Children's Hospital of Philadelphia, Philadelphia, Pennsylvania. napolitanon@email.chop.edu.
    • Respir Care. 2021 Oct 1; 66 (10): 151415201514-1520.

    BackgroundThe RAM cannula (Neotech, Valencia, CA) has become a commonly used interface for CPAP in neonatal intensive care. Performance characteristics of this interface used with a critical care ventilator are not well described.MethodsThis was a bench study utilizing a lung simulator configured as an actively breathing infant (weights of 800 g, 1.5 kg, and 3 kg) with moderate lung disease and a critical care ventilator in CPAP mode with leak compensation on. Three sizes of the RAM cannulae (preemie, newborn, and infant) were compared to 3 BabyFlow nasal prongs (Dräger Medical, Lübeck, Germany) (medium, large, and extra-large). Fabricated nasal models produced a 70% occlusive fit for the RAM cannula and an occlusive fit with the Dräger prongs. Delivered flow and pressure levels were recorded at 9 CPAP levels between 5 and 20 cm H2O.ResultsThe Dräger prongs produced a mean airway pressure ([Formula: see text]) within 0.20 cm H2O (range -0.10 to 0.35) of the set CPAP across all evaluated prong sizes and CPAP levels. In contrast, the RAM cannula produced [Formula: see text] values that averaged 8.5 cm H2O (range -15 to -3.5) below the set CPAP levels. The deficit in delivered versus target CPAP level for the RAM cannula increased with greater set CPAP. Set CPAP of 5 cm H2O delivered [Formula: see text] values that ranged from 0.6 to 1.5 cm H2O (difference of 3.5-4.4 cm H2O). Set CPAP of 20 cm H2O delivered [Formula: see text] values that ranged from 5.0 to 8.4 cm H2O (difference of 11.7-15 cm H2O). Inspiratory flow required to achieve set CPAP levels did not differ between interfaces, suggesting high resistance in the RAM cannula device masks the delivered CPAP levels.ConclusionsUse of the RAM cannula with a 30% leak on a critical care ventilator delivered [Formula: see text] values lower than set CPAP. This may be clinically meaningful and should be considered when choosing a nasal interface.Copyright © 2021 by Daedalus Enterprises.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.