    Analytical performance of time-resolved fluorometry-based Innotrac Aio! cardiac marker immunoassays.

    • P Hedberg, J Valkama, and M Puukka.
    • Laboratory of Oulu University Hospital, Department of Clinical Chemistry, University of Oulu, Finland. pirjo.hedberg@ppshp.fi
    • Scand. J. Clin. Lab. Invest. 2003 Jan 1; 63 (1): 55-64.

    AbstractThe results of an evaluation of the Innotrac Aio! cardiac markers are presented. This system is based on dry-chemistry, time-resolved fluorometry. All assay-specific reagents are dry-coated into assay-specific cups, and only the generic assay buffer is required. The levels of precision attained with pooled serum samples and control materials were acceptable for cTnI and CK-MB. Myoglobin assay showed higher CV, 5.6-9.5%. The linearity studies were performed in concentration ranges of 0.1-76 microg/L for cTnI, 0.7-450 microg/L for CK-MB and 0.6-1500 microg/L for myoglobin. The markers were found to be linear within the ranges tested. The correlation coefficient between the Aio! and AxSYM cTnI assays was 0.960, and the slope was 0.07. The correlation coefficients between the Aio! and AxSYM CK-MB and myoglobin assays were 0.995 and 0.971, respectively. They involved some differences in the measured concentrations (Aio! CK-MB was about 9% higher than AxSYM CK-MB, and Aio! myoglobin was 19% higher than AxSYM). Comparative studies with all the markers, using EDTA whole blood and lithium heparin plasma specimens and lithium heparin whole blood and plasma, yielded the following results: the slopes were close to 1.0 for all correlations, with the exception of that between CK-MB EDTA whole blood and lithium heparin (0.83). High correlation coefficients were obtained (> or = 0.97). The carryover results for all the cardiac markers were good, 0.0%, 0.0%, and 0.3% for cTnI, CK-MB, and myoglobin, respectively. The analytical detection limits were 0.01 microg/L for cTnI, 0.8 microg/L for CK-MB and 0.5 microg/L for myoglobin. The stability of the analytes in the lithium heparin samples at room temperature was also studied and was found to be decreased by from 10% (myoglobin and CK-MB) to 17% (cTnI) in 8 h. Innotrac Aio! provides a rapid and easy quantitative measurement of cardiac TnI, CK-MB, and myoglobin within < 18 min. This system is therefore suitable for use in emergency departments, coronary care units or central laboratory settings.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.