• Can J Anaesth · May 2004

    Comparative Study Clinical Trial

    Forehead is as sensitive as finger pulse oximetry during general anesthesia.

    • Shigekazu Sugino, Noriaki Kanaya, Masahito Mizuuchi, Masayasu Nakayama, and Akiyoshi Namiki.
    • Department of Anesthesiology, Sapporo Medical University School of Medicine, Sapporo, Japan.
    • Can J Anaesth. 2004 May 1;51(5):432-6.

    PurposeTo compare the performance of a forehead probe to a conventional finger pulse oximetry probe in anesthetized patients.MethodsEighteen patients participated in the study. Each probe was connected to a Nellcor N-550 pulse oximeter. Anesthesia was induced and maintained with propofol. After intubation, the patients received air to achieve a steady-state of peripheral arterial oxygen saturation (SpO(2)). Ventilation was interrupted to induce a hypoxic state. As soon as one of the two SpO(2)'s decreased to 90%, the patients' lungs were ventilated with 100% oxygen. To evaluate the performance of the two pulse oximeters, time to the lowest (TL), time of recovery (TR) and lag times to beginning of SpO(2) decrease (Lag) were measured.ResultsThere were no significant differences in TL and TR between forehead and finger pulse oximetry under normal perfusion conditions during general anesthesia. When the axillary artery was compressed to mimic reduced peripheral perfusion, SpO(2) in the forehead decreased sooner than in the finger during hypoxia. The forehead and finger TLs were similar, however, TR was significantly longer in the finger.ConclusionThe forehead SpO(2) sensor can be used as an alternative to the conventional finger sensor during general anesthesia.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.