    HypothesisOssicular implants made from metallic materials may be acceptable or pose hazards for patients referred for magnetic resonance imaging (MRI) examinations, depending on the outcome of proper MRI testing procedures.BackgroundUsing a 3-T MR system, 2 ossicular implants were tested for magnetic field interactions, heating, and artifacts.Materials And MethodsTwo different ossicular implants (Stainless Steel/Fluoroplastic Sanna-Type Piston [6 mm in length] and the Offset ALTO Total Prosthesis [15 mm in length, titanium/silicone]; Grace Medical, Memphis, TN, USA) were selected for testing, which represented the largest metallic mass and materials with the highest magnetic susceptibilities, with the intent of applying the MRI findings to other ossicular implants. The implants were evaluated at 3-T for magnetic field interactions, heating, and artifacts using standard previously described techniques.ResultsEach ossicular implant showed relatively minor magnetic field interactions that will not be associated with movement in situ. Heating was not excessive (highest temperature change, ≤ 1.6°C; background temperature change, 1.5°C). Artifacts, although relatively small, may create issues for diagnostic imaging if the area of interest is in the same area or close to these ossicular implants.ConclusionThe results of this investigation demonstrated that it would be acceptable (i.e., "MR conditional" using current terminology) for patients with these ossicular implants to undergo MRI examinations at 3 T or less. In consideration of the materials and dimensions of the implants that underwent testing, these findings pertain to many other similar ossicular implants from the same manufacturer.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…