• Clin Chem Lab Med · Jan 2006

    Comparative Study

    Comparability of point-of-care whole-blood electrolyte and substrate testing using a Stat Profile Critical Care Xpress analyzer and standard laboratory methods.

    • Zlata Flegar-Mestrić and Sonja Perkov.
    • Institute of Clinical Chemistry, Merkur University Hospital, Zagreb, Croatia. zlata.mestric@zh.htnet.hr
    • Clin Chem Lab Med. 2006 Jan 1; 44 (7): 898-903.

    BackgroundRapid technological progress in point-of-care testing allows the measurement of multiple analytes in whole-blood samples. The present study evaluated biosensor-based methods for the measurement of electrolytes and substrates in whole blood using a Stat Profile Critical Care Xpress (Nova Biomedical, Waltham, MA, USA) multiprofile analyzer and their comparability with standard laboratory methods. Because of the increased utilization of arterial blood samples in hospitalized patients and limited information on differences between arterial and venous blood for most routine laboratory tests, analytical differences caused by different sample types were evaluated.MethodsWhole-blood arterial samples and venous serum samples were obtained from 70 random patients with a variety of diagnoses admitted to the intensive care unit. The Stat Profile Critical Care Xpress analyzer was used to obtain whole-blood electrolyte and substrate profiles. For comparison studies, plasma or serum samples were analyzed according to standard laboratory methods using an Olympus AU 600 analyzer (Olympus Mishima, Shizuoka, Japan).ResultsImprecision, expressed as the coefficient of variation (CV%), was less than 5.7% for all analytes at both high and low concentrations, except for creatinine, with a CV of 13.8% for low and 9.5% for high concentrations. The inaccuracy of electrolyte and substrate measurements in whole blood using a Stat Profile Critical Care Xpress analyzer met the analytical quality specification required for near patient testing, with observed bias within the range -4.5% to 5.3%. Statistically significant correlation (p<0.05) was obtained between standard laboratory methods performed on arterial plasma or venous serum samples on an Olympus AU 600 analyzer and direct whole-blood measurements on the Stat Profile Critical Care Xpress point-of-care analyzer for all parameters tested, although slope and intercept values showed analytical differences for electrolyte measurement.ConclusionsThe Stat Profile Critical Care Xpress multiprofile point-of-care analyzer provides rapid and accurate direct whole-blood measurement with acceptable performance compared to standard laboratory methods. The results obtained for electrolytes and substrates in whole blood were comparable to those for standard laboratory methods using arterial plasma or venous serum samples.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.