    Practical application of biological variation and Sigma metrics quality models to evaluate 20 chemistry analytes on the Beckman Coulter AU680.

    • Mai Thi Chi Tran, KienTrung Hoang, and Ronda F Greaves.
    • Clinical Biochemistry Laboratory, National Hospital of Pediatrics, Hanoi, Viet Nam; Hanoi Medical University, Hanoi, Viet Nam. Electronic address: ungduongmai@gmail.com.
    • Clin. Biochem. 2016 Nov 1; 49 (16-17): 1259-1266.

    ObjectivesThis study aimed to evaluate the imprecision and bias data generated for 20 routine chemistry analytes against both the biological variation fitness for purpose (FFP) and Sigma metrics (SM) criteria.Design And MethodTwenty serum/plasma analytes were evaluated on the Beckman Coulter AU680. Third party commercial lyophilized internal quality control samples of human origin were used for day-to-day imprecision calculations. Commercial external quality assurance (EQA) samples were used to determine the systematic error between the test method result and the instrument group mean result from the EQA program for each analyte. Biological variation data was used to calculate the minimum, desirable and optimal imprecision and bias for determination of FFP. The desirable total allowable error was determined from biological variation data and applied to the SM calculation. The outcomes of both quality approaches were then compared.ResultsThe day-to-day imprecision of most tested analytes (except sodium and chloride) were smaller than the allowable imprecision (ranging from minimum to optimum). Most analytes achieved at least minimum bias. The SM varied with analyte concentration with six analytes producing low Sigma values. Comparing the quality processes eleven analytes produced a green light for both FFP and SM. There was some difference seen in interpretation for the other nine analytes.ConclusionsThe individual interpretation of bias and imprecision using FFP criteria allowed for the clear determination of the major source of error. Whereas, SM provided a summative evaluation of method performance. But the selection of total allowable error (TEa) is fundamental to this interpretation and harmonisation of the TEa calculation is needed.Copyright © 2016 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.