Randomized Controlled Trial Comparative Study
Development of a novel iPad-based laparoscopic trainer and comparison with a standard laparoscopic trainer for basic laparoscopic skills testing.
- Renai Yoon, Michael Del Junco, Adam Kaplan, Zhamshid Okhunov, Philip Bucur, Martin Hofmann, Reza Alipanah, Elspeth M McDougall, and Jaime Landman.
- Department of Urology, University of California, Irvine, California.
- J Surg Educ. 2015 Jan 1; 72 (1): 41-6.
IntroductionWe developed the iTrainer (iT) as a portable laparoscopic trainer, which incorporates the iPad tablet. We then compared the iT with a standard pelvic trainer (SPT) to assess surgical skills as well as its image quality, resolution, brightness, comfort, and overall performance.Materials And MethodsWe designed and constructed the iT to be compatible with the Apple iPad 3 and standard laparoscopic instruments. Participants were assigned to perform the thread-the-loops task on both trainers and were prospectively randomized to start on either the iT or the SPT. Each participant was allowed a 2-minute warm-up before the 2-minute testing period. We scored participants using the product of skill quality (0-4 scale) and quantity of loops threaded (0-10 scale). Participants then rated each trainer on image quality, resolution, brightness, comfort, and overall performance on a 5-point Likert scale.ResultsA total of 45 subjects including 10 undergraduates, 10 medical students, 10 general surgery and urology residents, and 15 experts (fellows and attending surgeons) participated in this study. There was no significant difference between thread-the-loops task scores completed on the iT when compared with the SPT for all groups tested (p > 0.05) with the exception of the medical student group, who performed better on the SPT (p < 0.05). On evaluation of each trainer, participants rated the iT as having superior image quality and resolution when compared with the SPT (p < 0.05) but rated the SPT higher in overall performance (p < 0.05). Brightness and comfort were rated similarly for both trainers.ConclusionsWe have demonstrated face validity and criterion validity for the thread-the-loops task on the iT. The iT rated superior in image quality and resolution but inferior in overall performance compared with the SPT. The iT provides trainees a unique advantage over SPT as an additional resource to laparoscopic training as it is inexpensive, portable, and can be readily available for training.Copyright © 2014 Association of Program Directors in Surgery. Published by Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.