Arch. Pathol. Lab. Med. · Jul 1996
Pneumatic transport exacerbates interference of room air contamination in blood gas samples.
- J R Astles, D Lubarsky, B Loun, F A Sedor, and J G Toffaletti.
- Department of Pathology, Division of Clinical Laboratories, Duke University Medical Center, Durham, NC 27710, USA.
- Arch. Pathol. Lab. Med. 1996 Jul 1; 120 (7): 642-7.
ObjectiveTo characterize and control the potential interference to P(O2) determinations when blood contaminated with air is sent via a pneumatic tube system (PTS).DesignBoth tonometered blood at P(O2)s of 65, 75, 142, and 339 mm Hg and arterial blood gas samples from patients with baseline P(O2)s from 70 to 400 mm Hg were analyzed for P(O2) to determine possible effects of air contamination from PTS transport.SettingA large teaching hospital in which a variety of personnel routinely send samples to the laboratory by PTS transport.PatientsTwenty patients under anesthesia for elective surgery and 21 patients in an intensive care unit who had a wide range of P(O2)s. Several additional patients with a preexisting lung pathology likely to cause hypoxemia were selected to provide samples with low P(O2)s.Main Outcome MeasuresMeasurement of bias in P(O2) between samples sent via PTS and samples walked to the laboratory.ResultsInterference from air contamination was worse after PTS transport compared with manual transport of the specimen. Over a wide range, the P(O2) in specimens after PTS transport tended toward 160 mm Hg. Samples from hypoxemic patients were prone to errors in P(O2) that could have resulted in clinical misinterpretation; 5 of 10 samples with a baseline P(O2) less than 85 mm Hg had increases of 10 mm Hg or more when contaminated with air. Cooling samples with high P(O2)s minimized changes to P(O2), probably by increasing the solubility of oxygen. Mechanical buffering by various liners used in the carriers did little to alleviate the interference. Decreasing the speed of pneumatic transport by 50% lessened the effect on P(O2).ConclusionInterference can be minimized by carefully purging samples of all air bubbles using the following protocol: invert syringe to check for air bubbles, then retap and reexpel bubbles if necessary. Personnel that collect and send blood gas samples via PTS should be educated about the problem of interference. Modifications both to pneumatic sample transport systems and to blood gas syringes should be investigated to minimize the effect.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.