Comparative Study
Optimization of keV-settings in abdominal and lower extremity dual-source dual-energy CT angiography determined with virtual monoenergetic imaging.
- Sonja Sudarski, Paul Apfaltrer, John W Nance, David Schneider, Mathias Meyer, Stefan O Schoenberg, Christian Fink, and Thomas Henzler.
- Institute of Clinical Radiology and Nuclear Medicine, University Medical Center Mannheim, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany. sonja_sudarski@gmx.de
- Eur J Radiol. 2013 Oct 1; 82 (10): e574-81.
ObjectivesTo compare objective image quality indices in dual-energy CT angiography (DE-CTA) studies of the abdomen and lower extremity using conventional polyenergetic images (PEIs) and virtual monoenergetic images (MEIs) at different kiloelectron volt (keV) levels.MethodsWe retrospectively evaluated 68 dual-source DE-CTA studies. 50 patients (42 men, 71 ± 10 years) underwent abdominal DE-CTA. 18 patients (13 men, 67 ± 10 years) underwent lower extremity DE-CTA. MEIs from 40 to 120 keV were reconstructed. Signal intensity, noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were assessed in infrarenal aorta, superior mesenteric, external iliac, femoral, popliteal, and lower leg arteries. Comparisons between MEIs and PEIs were performed with Dunnett's test.Results222 arteries were evaluated. In abdominal arteries 70 keV MEIs showed statistically equal signal intensity, noise and CNR levels (+13%; +31%, -14% on average; all p>0.05) compared to PEIs; SNR was equal or slightly impaired (-7% on average; p<0.001-1.00). In lower extremity arteries 60 keV MEIs resulted in significantly higher signal intensity and CNR (+54%; +54% on average; all p<0.05) compared to PEIs at equal noise levels (+18% on average; all p>0.05) and equal or higher SNR (+49% on average; p<0.01-0.35).ConclusionsLow-keV MEIs lead to equal or higher signal intensity and CNR compared to PEIs. In lower extremity DE-CTA, additional reconstruction of low-keV MEIs at 60 keV might increase diagnostic confidence.Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.