- Trevor R Born, Alison M Biercevicz, Sarath C Koruprolu, David Paller, Dave Spenciner, and Paul D Fadale.
- Brown University/Rhode Island Hospital, Providence, Rhode Island, U.S.A.. Electronic address: trevor_born@brown.edu.
- Arthroscopy. 2016 Feb 1; 32 (2): 253-61.
PurposeTo evaluate and compare two adjustable femoral cortical suspensory fixation devices used for anterior cruciate ligament reconstruction through a novel, direct computed tomography (CT) analysis metric and biomechanical laxity testing in a matched cadaveric human knee study.MethodsAnterior cruciate ligament reconstructions with bovine tendon grafts were performed using two adjustable femoral cortical suspensory fixation devices (RigidLoop Adjustable [DePuy Synthes Mitek, Raynham, MA] and TightRope [Arthrex, Naples, FL]) in 12 knees (6 matched pairs). A mechanical testing series was used to determine each knee's laxity in the intact condition. After reconstruction, each specimen was again tested for laxity and also imaged with CT. The laxity testing and CT imaging were then repeated after 1,000 cycles of anteroposterior loading on each knee to compare changes in laxity for the two fixation devices and to visualize changes in button-to-graft distance migration through a three-dimensional CT imaging method.ResultsNo significant differences were found between the two fixation groups' laxity measures after reconstruction (all P values ≥ .620) or after cycling (all P values ≥ .211) at any flexion angle. In addition, no significant differences were found between the two groups regarding button-to-graft distance migration (P = .773; mean, 0.61 ± 0.6 mm [95% confidence interval, -0.1 to 1.3 mm] in RigidLoop Adjustable group and 0.53 ± 0.6 mm [95% confidence interval, -0.1 to 1.2 mm] in TightRope group).ConclusionsThere were no significant differences between the two femoral cortical suspensory adjustable-loop devices regarding laxity outcomes or loop displacement as measured by button-to-graft distance migration.Clinical RelevanceUse of either of the adjustable-loop cortical suspensory devices in our analysis would appear to produce similar, acceptable laxity outcomes and minimal effects in terms of device-related loop displacement.Copyright © 2016 Arthroscopy Association of North America. Published by Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.