• Journal of hypertension · Jul 2001

    Comparative Study

    Ambulatory heart rate is underestimated when measured by an ambulatory blood pressure device.

    • T G Vrijkotte and E J de Geus.
    • Department of Biological Psychology, Vrije Universiteit, Amsterdam, The Netherlands. tgm.vrijkotte@psy.vu.nl
    • J. Hypertens. 2001 Jul 1; 19 (7): 1301-7.

    ObjectiveTo test the validity of ambulatory heart rate (HR) assessment with a cuff ambulatory blood pressure (ABP) monitor.DesignCross-instrument comparison of HR measured intermittently by a cuff ABP monitor (SpaceLabs, Redmond, Washington, USA), with HR derived from continuous electrocardiogram (ECG) recordings (1) in a controlled laboratory experiment and (2) during long-term recording in a true naturalistic setting.ParticipantsSix normotensive subjects participated in the laboratory study. A total of 109 male white-collar workers underwent ambulatory monitoring, of which 30 were mildly hypertensive.MethodsFour different laboratory conditions (postures: lying, sitting, standing, walking), repeated twice, were used to assess the short-term effects of cuff inflation on the HR. To test the actual ambulatory validity, participants simultaneously wore a continuous HR recorder and the ABP monitor from early morning to late evening on 2 workdays and one non-workday. Diary and vertical accelerometery information was used to obtain periods of fixed posture and (physical) activity across which HR from both devices was compared.ResultsLaboratory results showed that the ABP device reliably detected HR during blood pressure measurement, but that this HR was systematically lower than the HR directly before and after the blood pressure measurement. The ambulatory study confirmed this systematic underestimation of the ongoing HR, but additionally showed that its amount increased when subjects went from sitting to standing to light physical activity (2.9; 4.3 and 9.1 bpm (beats/min), respectively). In spite of this activity-dependent underestimation of HR, the correlation of continuous ECG and intermittent ABP-derived HR was high (median r= 0.81). Also, underestimation was not different for normotensives and mild hypertensives.ConclusionsA direct effect of cuff inflation leads to the underestimation of ongoing HR during cuff-based ABP measurement. Additional underestimation of HR occurs during periods with physical activity, probably due to behavioural freezing during blood pressure measurements. HR underestimation was not affected by hypertensive state. When its limitations are taken into account, ABP-derived ambulatory HR can be considered a reliable and valid measure.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.