- Teresa Ferrara, Barbara Baiotto, Giuseppe Malinverni, Nicola Caria, Elisabetta Garibaldi, Gilberto Barboni, Michele Stasi, and Pietro Gabriele.
- Radiooncology Department, A Businco Oncologic Hospital, Cagliari, Italy. dott.teresaferrara@yahoo.it
- Tumori. 2010 Jan 1; 96 (1): 76-83.
Aims And BackgroundA prospective analysis was carried out on a group of irradiated patients with pacemakers or implantable cardioverter-defibrillators to identify any relationship between the various types of devices and physical and dosimetric parameters. Cardiac toxicity of the treatment was also investigated.Methods And Study DesignForty-five irradiated patients, implanted with pacemakers or implantable cardioverter-defibrillators, were prospectively investigated from 1999 to 2007. An analysis of radiation damage to pacemakers, depending on the geometric and dosimetric characteristics of the radiation beams, was carried out. The electric and magnetic fields of linear accelerators (LINACs) were measured to evaluate any interference. The calculation of dose received by pacemakers was evaluated by dose-volume histograms.ResultsNo dysfunction was observed in any pacemaker, nor were the substitution times negatively affected. We did not find problems with the devices due to the interaction with the electromagnetic fields. Dose-volume histograms calculated for patients treated in regions close to devices (head & neck, thorax) showed an average maximum dose equal to 2.5 Gy for the head & neck area and equal to 1.8 Gy for the thoracic area. Acute (3 cases) and late (2 cases) cardiac events were observed only in 5 patients who underwent chemoradiation treatment, but no dysfunction was observed in any pacemaker.ConclusionsOur study confirms the safety of radiotherapy for patients implanted with pacemakers or implantable cardioverter-defibrillators but suggests that chemoradiation represents a probable risk factor for cardiac toxicity. Furthermore, all cardiac events were observed in patients treated in the head and'neck or left thoracic areas. A standardized protocol is advisable in order to improve patient control during the radiotherapy treatment. It is mandatory to calculate the dose received at the pacemaker/heart, even in the case of palliative treatment.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.