• Int. J. Radiat. Oncol. Biol. Phys. · Mar 1998

    Electronic portal imaging with on-line correction of setup error in thoracic irradiation: clinical evaluation.

    • J Van de Steene, F Van den Heuvel, A Bel, D Verellen, J De Mey, M Noppen, M De Beukeleer, and G Storme.
    • Department of Radiotherapy-Oncologic Center, Free University of Brussels, Jette, Belgium.
    • Int. J. Radiat. Oncol. Biol. Phys. 1998 Mar 1; 40 (4): 967-76.

    PurposeTo analyze setup errors and the feasibility of their on-line correction using electronic portal imaging in the irradiation of lung tumors.Methods And MaterialsSixteen patients with lung cancer were irradiated through opposed anteroposterior fields. Localization images of anteroposterior fields were recorded with an electronic portal imaging device (EPID). Using an in-house developed algorithm for on-line comparison of portal images setup errors were measured and a correction of table position was performed with a remote couch control prior to treatment. In addition, residual errors were measured on the EPID verification image. Global and individual mean and standard deviation of setup errors were calculated and compared. The feasibility of the procedure was assessed measuring intra- and interobserver variability, influence of organ movement, reproducibility of error measurement, the extra time fraction needed for measuring and adjusting and the fraction of dose needed for imaging.ResultsIn two setups the procedure could not be finished normally due to problems inherent to the procedure. The reproducibility, intraobserver variability, and influence of organ movements were each described by a distribution with a mean value less than or equal to 1 mm and a standard deviation (SD) of less than 1.5 mm. The interobserver variability showed to be a little bit larger (mean: 0.3 mm, SD: 1.7 mm). The mean time to perform the irradiation of the anteroposterior field was 4 +/- 1 min. The mean time for the measurement and correction procedure approximated 2.5 min. The mean extra time fraction was 65 +/- 24% (1 SD) with more than half of this coming from the error measurement. The dose needed for generation of EPID images was 5.9 +/- 1.4% of total treatment dose. The mean and SD of setup errors were, respectively, 0.1 and 4.5 mm for longitudinal and -2.0 and 5.7 mm for transversal errors. Of 196 measured translational errors 120 (61%) exceeded the adjustment criteria. For individual patients systematic and random setup errors can be as high as, respectively, 15.8 and 7.5 mm. Mean residual error and SD were for longitudinal direction 0.08 and 1.2 mm and for transversal direction -0.9 and 1.0 mm (pooled data). For individuals, the mean residual errors were smaller than 1 mm, with a typical SD per patient of less than 2 mm.ConclusionSetup errors in thoracic radiation therapy are clinically important. On-line correction can be performed accurately with an objective measurement tool, although this prolongs the irradiation procedure for one field with 65%.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…