• ASAIO J. · Jan 2004

    Characterization of an adult mock circulation for testing cardiac support devices.

    • George M Pantalos, Steven C Koenig, Kevin J Gillars, Guruprasad A Giridharan, and Dan L Ewert.
    • Jewish Hospital Heart and Lung Institute, Department of Surgery, University of Louisville, Kentucky 40202, USA. gmpant02@athena.louisville.edu
    • ASAIO J. 2004 Jan 1; 50 (1): 37-46.

    AbstractA need exists for a mock circulation that behaves in a physiologic manner for testing cardiac devices in normal and pathologic states. To address this need, an integrated mock cardiovascular system consisting of an atrium, ventricle, and systemic and coronary vasculature was developed specifically for testing ventricular assist devices (VADs). This test configuration enables atrial or ventricular apex inflow and aortic outflow cannulation connections. The objective of this study was to assess the ability of the mock ventricle to mimic the Frank-Starling response of normal, heart failure, and cardiac recovery conditions. The pressure-volume relationship of the mock ventricle was evaluated by varying ventricular volume over a wide range via atrial (preload) and aortic (afterload) occlusions. The input impedance of the mock vasculature was calculated using aortic pressure and flow measurements and also was used to estimate resistance, compliance, and inertial mechanical properties of the circulatory system. Results demonstrated that the mock ventricle pressure-volume loops and the end diastolic and end systolic pressure-volume relationships are representative of the Starling characteristics of the natural heart for each of the test conditions. The mock vasculature can be configured to mimic the input impedance and mechanical properties of native vasculature in the normal state. Although mock circulation testing systems cannot replace in vivo models, this configuration should be well suited for developing experimental protocols, testing device feedback control algorithms, investigating flow profiles, and training surgical staff on the operational procedures of cardiovascular devices.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…