• Magn Reson Med · Sep 2005

    Cramér-Rao bounds for three-point decomposition of water and fat.

    • Angel R Pineda, Scott B Reeder, Zhifei Wen, and Norbert J Pelc.
    • Department of Radiology, Stanford University Medical Center, Stanford, California 94305, USA. pineda@stanford.edu
    • Magn Reson Med. 2005 Sep 1; 54 (3): 625-35.

    AbstractThe noise analysis for three-point decomposition of water and fat was extended to account for the uncertainty in the field map. This generalization leads to a nonlinear estimation problem. The Crámer-Rao bound (CRB) was used to study the variance of the estimates of the magnitude, phase, and field map by computing the maximum effective number of signals averaged (NSA) for any choice of echo time shifts. The analysis shows that the noise properties of the reconstructed magnitude, phase, and field map depend not only on the choice of echo time shifts but also on the amount of fat and water in each voxel and their alignment at the echo. The choice of echo time shifts for spin-echo, spoiled gradient echo, and steady-state free precession imaging techniques were optimized using the CRB. The noise analysis for the magnitude explains rough interfaces seen clinically in the boundary of fat and water with source images obtained symmetrically about the spin-echo. It also provides a solution by choosing appropriate echo time shifts (-pi/6+pik, pi/2+pik, 7pi/6+pik), with k an integer. With this choice of echo time shifts it is possible to achieve the maximum NSA uniformly across all fat:water ratios. The optimization is also carried out for the estimation of phase and field map. These theoretical results were verified using Monte Carlo simulations with a newly developed nonlinear least-squares reconstruction algorithm that achieves the CRB.Copyright (c) 2005 Wiley-Liss, Inc.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.