• J. Heart Lung Transplant. · Dec 2015

    Left ventricular dimension decrement index early after axial flow assist device implantation: A novel risk marker for late pump thrombosis.

    • Emer Joyce, Garrick C Stewart, Maryclare Hickey, Lara Coakley, Jose M Rivero, Igor Gosev, Michael M Givertz, Gregory S Couper, and Mandeep R Mehra.
    • Center for Advanced Heart Disease, Brigham and Women's Hospital Heart and Vascular Center, Boston, Massachusetts; Harvard Medical School, Boston, Massachusetts.
    • J. Heart Lung Transplant. 2015 Dec 1; 34 (12): 1561-9.

    BackgroundIt is increasingly recognized that pump thrombosis most likely represents the end stage of a complex interaction between the patient-pump interface. We hypothesized that early patient/pump mismatch, as manifested by suboptimal left ventricular (LV) unloading early after left ventricular assist device (LVAD) implantation, may be a harbinger of increased risk for later LVAD thrombosis.MethodsIn 64 patients (59 ± 11 years old, 78% men, 44% destination therapy) discharged alive without thrombosis or other device malfunction after first HeartMate II LVAD implantation (between January 2011 and June 2014), LV dimensions in end diastole (LVIDd) and end systole (LVIDs) were compared between pre-implant and optimal set speed pre-discharge echocardiography. LV dimension decrement indices (pre-implant dimension - optimal set speed dimension ÷ pre-implant dimension × 100) for LVIDd [LVIDdDI] and LVIDs [LVIDsDI] were calculated.ResultsThe incidence of pump thrombosis was 0.06 per patient year (n = 18, median time 8 [interquartile range 2, 17] months). Baseline characteristics including pre-operative LVIDd and LVIDs were similar between LVAD thrombosis and no thrombosis groups. After ventricular assist device implantation, set speed and other ramp parameters did not differ between groups. However, LVIDdDI (19 ± 13% vs 25 ± 11%, p = 0.04) and LVIDsDI (16 ± 16% vs 27 ± 13%, p = 0.008) were significantly lower in patients with later pump thrombosis. A cutoff value of ≤15% using receiver operating characteristic curve analysis was 83% sensitive for LVIDdDI and LVIDsDI for predicting pump thrombosis. Patients with LVIDdDI of >15% vs ≤15% were significantly more likely to be free of pump thrombosis over a median follow-up period of 15 (interquartile range 9, 26) months (log-rank test, p = 0.045).ConclusionsLV dimension decrement index at optimized speed setting on pre-discharge echocardiography is associated with LVAD thrombosis.Copyright © 2015 International Society for Heart and Lung Transplantation. Published by Elsevier Inc. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…