Randomized Controlled Trial Comparative Study Clinical TrialMechanical ventilators optimized for pediatric use decrease work of breathing and oxygen consumption during pressure-support ventilation.
- M F el-Khatib, R L Chatburn, D L Potts, J L Blumer, and P G Smith.
- Division of Pediatric Pharmacology, Rainbow Babies and Childrens Hospital, Cleveland, OH 44106.
- Crit. Care Med. 1994 Dec 1;22(12):1942-8.
Objectivesa) To investigate whether the patient work of breathing needed to trigger inspiration is affected by the type of ventilator delivering pressure-support ventilation for mechanically ventilated pediatric patients. b) To determine whether changes in oxygen consumption (VO2) trend with changes in work of breathing and would thus be helpful in tracking work of breathing.DesignProspective study.SettingPediatric intensive care unit at a university hospital.PatientsNine mechanically ventilated patients (2 to 75 months of age).InterventionsWhile maintaining a constant pressure-support ventilation level, patients were alternately supported with the Siemens Servo 900C, the Bird VIP, and the Newport Wave E200 ventilators in random order.Measurements And Main ResultsWork of breathing, defined as the integral of the pressure-volume curve corresponding to negative pressure, was calculated with a pulmonary monitoring system. VO2 was measured with a metabolic cart. Patient distress levels were assessed using the COMFORT scale, a behavioral scoring system. Mean values (20 breaths/patient) for measured variables with each ventilator were compared using analysis of variance and Scheffé tests, with p < .05 indicating statistical significance. The lowest VO2 (103 +/- 35 mL/min/m2) and work of breathing (24 +/- 15 g.cm/m2) were achieved with the Bird VIP ventilator and were significantly (p < .05) lower than those values obtained with either the Siemens Servo 900C (VO2 147 +/- 33 mL/min/m2; work of breathing 49 +/- 18 g.cm/m2) or the Newport Wave E200 (VO2 122 +/- 33 mL/min/m2; work of breathing 35 +/- 15 g.cm/m2). Also, the values of work of breathing and VO2 obtained using the Newport Wave E200 were significantly (p < .05) lower than those values obtained using the Servo 900C. No change in behavioral distress occurred when the ventilators were changed. In all patients, there was a clear similarity in the trends of VO2 and work of breathing.ConclusionsWe conclude that VO2 and work of breathing may be reduced significantly using the latest generation of mechanical ventilators optimized for infant and pediatric use. Because work of breathing is less with the Bird VIP than the other two ventilators tested, leading to a corresponding decrease in VO2, we suggest that the Bird VIP better adapts the patient to the ventilator and may facilitate weaning from ventilatory support. We also suggest that changes in VO2 might be helpful in tracking changes in work of breathing.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.