Clinical evaluation of the FloTrac/Vigileo system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery.
- D Button, L Weibel, O Reuthebuch, M Genoni, A Zollinger, and C K Hofer.
- Institute of Anaesthesiology and Intensive Care Medicine, Triemli City Hospital Zurich, Zurich, Switzerland.
- Br J Anaesth. 2007 Sep 1;99(3):329-36.
BackgroundAssessment of cardiac output (CO) by the FloTrac/Vigileo system may offer a less invasive means of determining the CO than either the pulmonary artery catheter (PAC) or the PiCCOplus system. The aim of this study was to compare CO measurements made using the FloTrac/Vigileo system with upgraded software (FCO, Edwards Lifesciences, Irvine CA, USA), the PiCCOplus system (PCO, Pulsion Medical Systems, Munich, Germany) and continuous CO monitoring using a PAC (CCO; Vigilance monitoring, Edwards Lifesciences, Irvine CA, USA) with intermittent pulmonary artery thermodilution (ICO). The study was conducted in patients undergoing elective cardiac surgery.MethodsThirty-one patients with preserved left ventricular function were enrolled. CCO, FCO, and PCO were recorded in the perioperative period at six predefined time points after achieving stable haemodynamic conditions; ICO was determined from the mean of three bolus injections. Bland-Altman analysis was used to compare CCO, FCO, and PCO with ICO.ResultsBland-Altman analysis revealed a comparable mean bias and limits of agreement for all tested continuous CO monitoring devices using ICO as reference method. Agreement for all devices decreased in the postoperative period.ConclusionThe performance of the FloTrac/Vigileo system, the PiCCOplus, and the Vigilance CCO monitoring for CO measurement were comparable when tested against intermittent thermodilution in patients undergoing elective cardiac surgery.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.