• Investigative radiology · Aug 2013

    Comparative Study

    Automated low-kilovoltage selection in pediatric computed tomography angiography: phantom study evaluating effects on radiation dose and image quality.

    • Marilyn Joy Siegel, Juan Carlos Ramirez-Giraldo, Charles Hildebolt, David Bradley, and Bernhard Schmidt.
    • Mallinckrodt Institute of Radiology, Washington University School of Medicine, St Louis, MO 63110, USA. siegelm@mir.wustl.edu
    • Invest Radiol. 2013 Aug 1; 48 (8): 584-9.

    PurposeThe purpose of this study was to compare the effects of combined automated tube voltage selection and automated tube current modulation on radiation dose and image quality in small-sized phantoms undergoing computed tomography angiography (CTA) with the use of automated current modulation alone.Materials And MethodsThree semianthropomorphic phantoms, corresponding to a neonate, a small child, and a small adult, underwent simulated abdominal CTA using an automated tube voltage selection technology, which simultaneously optimizes kilovoltage (in kilovolt [peak]) and tube-current/milliamperage (in milliampere-second) on the basis of the patient topogram and clinical task. The phantoms were scanned with 2 protocols: protocol A, using the combination of automated kilovoltage and milliamperage, and protocol B, using only automated milliamperage with the standard 120 kV(p). Radiation doses were measured for each phantom, and the measurements were then used to estimate volume computed tomography dose index. Image noise and iodine contrast, contrast-to-noise ratio, and the relative dose factor were assessed. Differences were tested using paired t tests, and percentage differences for various technical factors and the phantom sizes were calculated.ResultsThe computed tomography dose index was significantly lower in protocol A (mean, 3.3 mGy) compared with that in protocol B (mean, 7.7 mGy), representing a 56.0% dose reduction (P = 0.01). In protocol A, tube potential dropped from 120 to 70 kV(p) in the small and medium phantoms and to 80 kV(p) in the large phantom. For each phantom size, image noise and iodine contrast increased significantly in protocol A relative to those in protocol B (P = 0.03 and P < 0.01, respectively). Corresponding contrast-to-noise ratio values increased by 9.1% in protocol A relative to those in protocol B (P = 0.04). The relative dose factor values for protocol A relative to those for protocol B were 31%, 36%, and 44% for the small, medium, and large phantoms, respectively.ConclusionsCombined use of automated kilovoltage selection and automated tube current modulation is more effective for reducing radiation dose and maintaining image quality during simulated pediatric CTA than is automated tube current modulation in isolation.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…