Expert Rev Med Devices · Feb 2021
Reflection efficiencies of AnaConDa-S and AnaConDa-100 for isoflurane under dry laboratory and simulated clinical conditions: a bench study using a test lung.
- Azzeddine Kermad, Jacques Speltz, Philipp Daume, Thomas Volk, and Andreas Meiser.
- Department of Anesthesiology, Intensive Care Medicine and Pain Medicine, Saarland University Hospital Medical Center, Homburg, Germany.
- Expert Rev Med Devices. 2021 Feb 1; 18 (2): 189-195.
AbstractBackground: Adequate sedation is important for the treatment of ICU patients. AnaConDa (Anesthetic-Conserving-Device; ACD; Sedana Medical, Sweden), connected between ventilator and the patient, retains isoflurane during expiration, and releases it back during inspiration. The reflection efficiency (RefEff) corresponds to the percentage of expired anesthetic molecules that are re-inspired. We compared RefEff of AnaConDa-S (ACD-50) and AnaConDa-100 (ACD-100) under laboratory (DRY) and simulated clinical conditions (CLIN) using a test lung.Methods: Measurements were made under DRY and CLIN, with different tidal volumes (TV: 300 mL & 500 mL) and infusion rates (0.5-10 mL·h-1). RefEff was calculated from the isoflurane concentration in the test-lung (CISO) and plotted against the anesthetic vapor volume exhaled in one breath (V-exh = CISO·TV).Results: DRY: RefEff of both devices was ≈90% over a wide range of V-exh, but decreased when V-exh exceeded 5-7 mL (ACD-50) or 10-15 mL (ACD-100).CLIN: RefEff of ACD-50 was 70-80% (ACD-100: 80-90%), decreasing gradually with increasing V-exh. For 1 Vol.% isoflurane at TV500, the infusion rate with ACD-50 was twofold higher compared to ACD-100 (4 versus 2 mL·h-1).Conclusion: Under DRY and concentrations <1.5 Vol.%, RefEff of both devices is around 90%. Under CLIN, ACD-100 performs better with RefEff between 80% and 90% (ACD-50:70-80%), decreasing with increased vapor volume exhaled in one breath.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.