    Comparative Study

    Systematic Error of Cardiac Output Measured by Bolus Thermodilution With a Pulmonary Artery Catheter Compared With That Measured by an Aortic Flow Probe in a Pig Model.

    • Dewi K Rowlands and Zhu Fang.
    • Department of Intensive Care, Cancer Center, Sun Yat-sen University, Canton, China.
    • J. Cardiothorac. Vasc. Anesth.. 2013 Dec 1;27(6):1133-9.

    ObjectiveThe precision of thermodilution cardiac output measurement using a pulmonary artery catheter can be divided into random and systematic errors. This study determined the systematic component.DesignComparative validation against a transonic flow probe on the aortic root.SettingAnimal research laboratory.ParticipantsEight anesthetized pigs, weight 27 to 32 kg.InterventionsThermodilution measurements were compared to those from an aortic flow probe. One (or two) catheters were tested in each pig, with multiple paired readings recorded as cardiac output was varied pharmacologically by esmolol, epinephrine, or dopamine. Linear regression lines were drawn for each pig as well as the slope used to quantify systematic error. Regression analysis of data from each pig was used to assess trending. Bland-Altman analysis also was performed.Measurements And Main ResultsSystematic error derived from slope data was ± 26%. Trending was reliable in 8 out of 10 catheter placements (p value>0.95). Bland- Altman analysis (n = 77 basal anesthesia and 165 pharmacologic intervention data pairs) provided percentage errors of ± 23% and ± 34-39%, respectively. This percentage error increase was due to variations in calibration and the slopes of regression lines for each catheter tested as the lines diverged as cardiac output increased, widening the spread of data.ConclusionsThermodilution does trend cardiac output reliably in most cases. However, the systematic error is large and has a significant effect on the percentage error as cardiac output increases. The precision error of ± 20% currently used for thermodilution measurement may be set too low and is dependent on the clinical setting.Copyright © 2013 Elsevier Inc. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…