- Árpád Farkas, David Lewis, Tanya Church, Alan Tweedie, Francesca Mason, Allen E Haddrell, Jonathan P Reid, Alpár Horváth, and Imre Balásházy.
- Centre for Energy Research, Hungarian Academy of Sciences, Konkoly-Thege Miklós út 29-33, 1121, Budapest, Hungary. Electronic address: farkas.arpad@energia.mta.hu.
- Int J Pharm. 2017 Nov 25; 533 (1): 225-235.
AbstractThe breath-actuated mechanism (BAM) is a mechanical unit included in NEXThaler® with the role of delaying the emission of the drug until the inhalation flow rate of the patient is sufficiently high to detach the drug particles from their carriers. The main objective of this work was to analyse the effect of the presence of BAM on the size distribution of the emitted drug and its airway deposition efficiency and distribution. Study of the hygroscopic growth of the emitted drug particles and its effect on the deposition was another goal of this study. Size distributions of Foster® NEXThaler® drug particles emitted by dry powder inhalers with and without BAM have been measured by a Next Generation Impactor. Three characteristic inhalation profiles of asthmatic patients (low, moderate and high flow rates) were used for both experimental and modelling purposes. Particle hygroscopic growth was determined by a new method, where experimental measurements are combined with simulations. Upper airway and lung deposition fractions were computed assuming 5s and 10s breath-hold times. By the inclusion of BAM the fine particle fraction of the steroid component increased from 24 to 30% to 47-51%, while that of bronchodilator from 25-34% to 52-55%. The predicted upper airway steroid and bronchodilator doses decreased from about 60% to 35-40% due to BAM. At the same time, predicted lung doses increased from about 20%-35% (steroid) and from 22% to 38% (bronchodilator) for the moderate flow profile and from about 25% to 40% (steroid) and from 29% to 47% (bronchodilator) for the high inhalation flow profile. Although BDP and FF upper airway doses decreased by a factor of about two when BAM was present, lung doses of both components were about the same in the BAM and no-BAM configurations at the weakest flow profile. However, lung dose increased by 2-3% even for this profile when hygroscopic growth was taken into account. In conclusion, the NEXThaler® BAM mechanism is a unique feature enabling high emitted fine particle fraction and enhanced drug delivery to the lungs.Copyright © 2017 The Authors. Published by Elsevier B.V. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.