Int. J. Radiat. Oncol. Biol. Phys. · Nov 2012
Evaluation of rotational errors in treatment setup of stereotactic body radiation therapy of liver cancer.
- Minsong Cao, Foster D Lasley, Indra J Das, Colleen M Desrosiers, Eric D Slessinger, and Higinia R Cardenes.
- Department of Radiation Oncology, Indiana University School of Medicine, Indianapolis, Indiana, USA.
- Int. J. Radiat. Oncol. Biol. Phys. 2012 Nov 1; 84 (3): e435-40.
PurposeTo evaluate the dosimetric impact of rotational setup errors in stereotactic body radiotherapy (SBRT) treatment of liver tumors and to investigate whether translational shifts can compensate for rotation.Methods And MaterialsThe positioning accuracy in 20 patients with liver malignancies treated with SBRT was reevaluated offline by matching the patients' cone-beam computed tomography (CT) scans (n=75) to the planning CT scans and adjusting the 3 rotational angles (pitch, roll, and yaw). Systematic and random setup errors were calculated. The dosimetric changes caused by rotational setup errors were quantified for both simulated and observed patient rotations. Dose distributions recalculated on the rotated CT scans were compared with the original planned doses. Translational corrections were simulated based on manual translational registration of the rotated images to the original CT scans. The correction efficacy was evaluated by comparing the recalculated plans with the original plans.ResultsThe systematic rotational setup errors were -0.06° ± 0.68°, -0.29° ± 0.62°, and -0.24° ± 0.61°; the random setup errors were 0.80°, 1.05°, and 0.61° for pitch, roll, and yaw, respectively. Analysis of CBCT images showed that 56.0%, 14.7%, and 1.3% of treated fractions had rotational errors of >1°, >2°, and >3°, respectively, in any one of the rotational axes. Rotational simulations demonstrated that the reduction of gross tumor volume (GTV) coverage was <2% when rotation was <3°. Recalculated plans using actual patient roll motions showed similar reduction (<2%) in GTV coverage. Translational corrections improved the GTV coverage to within 3% of the original values. For organs at risk (OAR), the dosimetric impact varied case by case.ConclusionActual rotational setup errors in SBRT for liver tumors are relatively small in magnitude and are unlikely to affect GTV coverage significantly. Translational corrections can be optimized to compensate for rotational setup errors. However, caution regarding possible dose increases to OAR needs to be exercised.Copyright © 2012 Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.