- Brian Fissel, Berton R Moed, and J Gary Bledsoe.
- Department of Orthopaedic Surgery, Saint Louis University School of Medicine, St Louis, MO 63103, USA.
- J Orthop Trauma. 2008 May 1; 22 (5): 337-41.
ObjectiveThe objective of this study was to test the stiffness and ultimate load to failure of new intramedullary (IM) nail proximal screw configurations as compared to a trochanteric reconstruction nail.MethodsTwenty-one synthetic composite femurs were mounted on a Material Testing System and tested in axial compression 5 times. The femurs had an 1 of 2 IM nail types inserted with 1 of 3 proximal screw configurations: a 3-screw configuration with 2 transverse screws and a screw angled into the femoral neck; a 2-screw design with a single transverse screw and a single screw angled into the femoral neck; 2 parallel screws angled into the femoral neck. There were 7 specimens in each group. An unstable fracture (OTA/AO 32-C3.2) was created. and the stiffness of these constructs was tested in compression 5 times. Each construct was then loaded to failure in compression.ResultsThe 3-screw construct provided more axial stiffness (214 N/mm +/- 75) than either the 2-screw construct (123 N/mm +/- 32) or the trochanteric reconstruction nail (127 N/mm +/- 21) (P = 0.017 and 0.035 for 3-screw vs. 2-screw and recon respectively, P = 0.45 for 2-screw vs. recon). Load-to-failure testing demonstrated similarity among the different screw configurations (3-screw = 2230 N +/- 265, 2-screw = 2283 N +/- 260, Reconstruction nail = 2121 N +/- 156) (P = 1.0 all groups).ConclusionsThe proximal 3-screw configuration provided more stiffness than either the 2-screw configuration or trochanteric reconstruction nail. The 2-screw configuration performed equally to a standard trochanteric reconstruction nail in stiffness testing. The ultimate loads to failure for the 3 tested constructs were not significantly different.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.