Randomized Controlled Trial
Comparison of the Efficiency and Usability of Aerosol Box and Intubation Tent on Intubation of a Manikin Using Personal Protective Equipment: A Randomized Crossover Study.
- Kin Wa Wong, LamRex Pui KinRPKEmergency Medicine Unit, Li Ka Shing Faculty of Medicine, The University of Hong Kong, Hong Kong Special Administrative Region, China; 24-hour Outpatient and Emergency Department, Gleneagles Hospital Hong Kong, Hong Kong Special Administr, Wai Ching Sin, Michael G Irwin, and Timothy H Rainer.
- Emergency Medicine Unit, Li Ka Shing Faculty of Medicine, The University of Hong Kong, Hong Kong Special Administrative Region, China; 24-hour Outpatient and Emergency Department, Gleneagles Hospital Hong Kong, Hong Kong Special Administrative Region, China.
- J Emerg Med. 2021 Dec 1; 61 (6): 695704695-704.
BackgroundThe aerosol box and intubation tent are improvised barrier-enclosure devices developed during the novel coronavirus pandemic to protect health care workers from aerosol transmission.ObjectiveUsing time to intubation as a crude proxy, we aimed to compare the efficiency and usability of the aerosol box and intubation tent in a simulated manikin.MethodsThis was a single-center, randomized, crossover manikin study involving 28 participants (9 anesthetists, 16 emergency physicians, and 3 intensivists). Each participant performed rapid sequence intubations in a random sequence of three different scenarios: 1) no device use; 2) aerosol box; 3) intubation tent. We compared the time to intubation between different scenarios.ResultsThe median total intubation time with no device use, aerosol box, and intubation tent were 23.7 s (interquartile range [IQR] 19.4-28.4 s), 30.9 s (IQR 24.1-52.5 s), and 26.0 s (IQR 22.1-30.8 s), respectively. Post hoc analysis showed a significantly longer intubation time using the aerosol box compared with no device use (p < 0.001) and compared with the intubation tent (p < 0.001). The difference between the intubation tent and no device use was not significant. The first-pass intubation success rate did not differ between the groups. Only aerosol box use had resulted in breaches of personal protective equipment. Participants considered intubation with the intubation tent more favorable than the aerosol box.ConclusionsThe intubation tent seems to have a better barrier-enclosure design than the aerosol box, with a reasonable balance between efficiency and usability. Further evaluation of its efficacy in preventing aerosol dispersal and in human studies are warranted prior to recommendation of widespread adoption.Copyright © 2021 Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.