- Mathieu Delorme, Pierre-Alexandre Bouchard, Serge Simard, and François Lellouche.
- Centre de Recherche de l'Institut Universitaire de Cardiologie et de Pneumologie de Québec, Quebec, Canada. Mr Delorme is affiliated with the Université Paris-Saclay, UVSQ, ERPHAN,Versailles, France.
- Respir Care. 2021 Nov 1; 66 (11): 172017281720-1728.
BackgroundHigh-flow nasal cannula (HFNC) is increasingly used for the management of respiratory failure. Settings include [Formula: see text], total gas flow, and temperature target. Resulting absolute humidity (AH) at the nasal cannula may affect clinical tolerance, and optimal settings with respect to hygrometry remain poorly documented.MethodsA bench study was designed to assess AH delivered by 4 HFNC devices (Optiflow, Airvo 2, Precision Flow, and Hydrate) according to flow, ambient temperature, and other available settings. Clinical tolerance of different levels of hygrometry (20, 30, and 40 mg H2O/L) was evaluated in 15 healthy volunteers.ResultsWith [Formula: see text] set at 1.0, normal ambient temperature, and settings made accordingly to the manufacturers' recommendations, mean ± SD AH was 42.2 ± 3.1, 39.5 ± 1.8, 35.7 ± 2.0, and 32.9 ± 2.7 mg H2O/L for the Airvo 2, Optiflow, Hydrate, and Precision Flow, respectively, (P < .001). AH dropped from -3.5 to -10.7 mg H2O/L (P <. 001) with high ambient temperature, except for the Precision Flow. Increasing flow did not significantly affect AH except for the Precision Flow (from 36.4 ± 1.6 to 29.8 ± 0.2 mg H2O/L at 10 and 40 L/min, respectively, [P < .001]). The lowest AH was encountered with the Optiflow set with noninvasive ventilation (NIV) mode, without compensation algorithm, and at high ambient temperature (14.2 ± 1.5 mg H2O/L). In studied subjects, AH significantly affected breathing comfort, reduced from 7.0 ± 1.0 to 3.0 ± 2.0 at 40 and 20 mg H2O/L, respectively, (P < .001). Comfort was similar at 30 and 40 mg H2O/L.ConclusionsWhen used according to manufacturer's recommendations and at normal ambient temperature, all the HFNC devices evaluated achieved satisfactory hygrometric output with respect to breathing comfort evaluated in healthy subjects (≥ 30 mg H2O/L). Substantial differences exist between devices, and optimal knowledge of their working principles is required as inappropriate usage may dramatically alter efficacy and clinical tolerance.Copyright © 2021 by Daedalus Enterprises.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.