Ann. Clin. Biochem. · Sep 2008
Case ReportsGlucose interference in direct ion-sensitive electrode sodium measurements.
- Asila Al-Musheifri and Graham R D Jones.
- Department of Chemical Pathology, St Vincent's Hospital, Victoria St Darlinghurst, Sydney, NSW 2010, Australia.
- Ann. Clin. Biochem. 2008 Sep 1; 45 (Pt 5): 530-2.
AbstractCirculating sodium concentration is commonly measured by both direct and indirect ion-sensitive electrode (ISE). We describe an unusual case with a high elevation of serum glucose (162 mmol/L) where direct ISE sodium measurement was 9 mmol/L higher than the indirect measurement in the absence of any cause for pseudohyponatraemia. In vitro experiments showed that very high glucose concentrations increased the sodium in direct, but not in indirect ISE measurement. This effect was insufficient to account for the entire difference between the measurements seen in the patient, indicating that other factors, for example pH and bicarbonate concentration, must also be involved. This effect may influence interpretation of sodium status in patients with gross hyperglycaemia.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.