Observational StudyAccuracy of noninvasive blood pressure measured at the ankle during cesarean delivery under spinal anesthesia.
- Sherin Refaat, Maha Mostafa, Ahmed Hasanin, Nadeem Rujubali, Reham Fouad, and Yasmin Hassabelnaby.
- Department of Anesthesia and Critical Care Medicine, Cairo University, Cairo, Egypt.
- J Clin Monit Comput. 2021 Oct 1; 35 (5): 1211-1218.
AbstractIn this study, we evaluated the accuracy of oscillometric noninvasive blood pressure (NIBP) measured at the ankle in detecting low arm NIBP during cesarean delivery under spinal anesthesia. In this prospective observational study, a cohort of full-term mothers undergoing elective cesarean delivery under spinal anesthesia was examined. Simultaneous NIBP measurements were obtained from the arm and the ankle. The primary outcome was the accuracy of the ankle NIBP in detecting arm systolic blood pressure (SBP) < 90 mmHg. Other outcomes included the accuracy of ankle NIBP in detecting SBP < 80% of the baseline value. The area under the receiver operating characteristic curve (AUC) was calculated to evaluate the accuracy of ankle NIBP in detecting low arm NIBP. The Bland-Altman analysis was conducted to evaluate the agreement between values. We analyzed 1729 pairs of readings obtained from 97 mothers. Ankle SBP showed good accuracy in detecting SBP < 90 mmHg, with an AUC (95% confidence interval [CI]) of 0.90 (0.89-0.91) and a negative predictive value (NPV) of 99 (98-99%) at a cutoff value of ≤ 103 mmHg. Furthermore, ankle SBP showed good accuracy in detecting SBP < 80% of the baseline value, with an AUC (95% CI) of 0.84 (0.82-0.89) and an NPV of 95 (93-96%) at a cutoff value of ≤ 76% of the ankle baseline SBP. The mean bias between the two sites of measurement was - 5.4 ± 15.5, - 2.0 ± 11, and 0.5 ± 12.1 mmHg for SBP, diastolic blood pressure, and mean arterial pressure, respectively. In conclusion, ankle NIBP measurement is not interchangeable with arm NIBP measurement. However, ankle NIBP measurement showed good accuracy for ruling out low arm NIBP during a cesarean delivery.Clinical trial rejistration: NCT04199156.© 2020. Springer Nature B.V.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.