• Health physics · Jun 2008

    Blind testing of commercially available short-term radon detectors.

    • Kainan Sun, Gregory Budd, Steven McLemore, and R William Field.
    • Department of Occupational and Environmental Health, Department of Biostatistics, College of Public Health, University of Iowa, Iowa City, IA 52242, USA.
    • Health Phys. 2008 Jun 1;94(6):548-57.

    AbstractA sample of commercially available, charcoal adsorption type, short-term radon detectors was blind tested under controlled laboratory conditions in order to obtain a "snapshot" of the accuracy and precision of the detectors. The results of the controlled exposures were then compared to a previous field study of the same type of commercially available radon detectors. Radon detectors, purchased from seven different commercial vendors, were exposed to a reference (222)Rn gas concentration at the U.S. Environmental Protection Agency's (EPA) Radon Chamber located at the Radiation and Indoor Environments National Laboratory in Las Vegas, Nevada. EPA Test 1 was performed under a controlled simulated field exposure paralleling, to the extent possible, the previous actual field exposure conditions. A second controlled exposure, EPA Test 2, was performed under a relatively steady state of (222)Rn gas concentration, at the same temperature, but a more moderate relative humidity. In the previous field setting evaluation of detectors, five out of six companies tested did not pass the accuracy guideline (all individual relative errors < or =25%) established during the EPA's former Radon Measurement Proficiency Program (EPA-RMPP). As compared to the field test, the detectors in this study generally exhibited better accuracy and precision. Not surprisingly, it appeared temporal fluctuations in radon concentrations and increased humidity had a negative influence on the accuracy and precision of detectors for some companies. The inability of three out of seven companies to meet former EPA-RMPP guidelines for accuracy, even under ideal exposure conditions (constant temperature, humidity, and radon concentration), highlights the importance of blind testing commercially available radon detectors. Furthermore, the consistent over-reporting or under-reporting trends in the overall results for all three tests suggest a potentially widespread systematic bias for the individual companies that merits further investigation. It is unknown if this one-time "snapshot" represents the overall reliability of commercially available charcoal-based radon detectors. Nonetheless, the findings suggest the need for improved vigilance to assure that the public can rely on commercially available radon detectors to make an informed decision whether or not to perform additional testing or to mitigate.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…