Case ReportsResolution of T-wave oversensing with implantable cardioverter defibrillator generator replacement in an adolescent.
- Volkan Tuzcu.
- Arkansas Children's Hospital, University of Arkansas for Medical Sciences, Little Rock, Arkansas 72202, USA. tuzcuvolkan@uams.edu
- Pacing Clin Electrophysiol. 2007 Jul 1; 30 (7): 929-32.
AbstractInappropriate delivery of therapy due to T-wave oversensing is a frequent complication of intracardiac cardioverter defibrillators (ICD). A 16-year-old adolescent with congenital long QT syndrome type 3 underwent single-chamber implantable cardioverter defibrillator (ICD) (EnTrust VR, Medtronic Inc., Minneapolis, MN, USA) implantation.He received two appropriate and four inappropriate shocks. The inappropriate shocks were due to T-wave oversensing. Following the replacement of ICD with a different brand generator (Vitality 2 VR, Boston Scientific Inc., Natick, MA, USA), no more T-wave oversensing is noted. Proper filtering is imperative in ICDs in order to prevent T-wave oversensing-related inappropriate delivery of therapy.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.