• J Clin Monit · Jul 1996

    Comparative Study

    Accuracy and precision of a new, portable, handheld blood gas analyzer, the IRMA.

    • J A Wahr, W Lau, K K Tremper, L Hallock, and K Smith.
    • Department of Anesthesiology, University of Michigan Medical School, Ann Arbor, USA.
    • J Clin Monit. 1996 Jul 1; 12 (4): 317-24.

    ObjectiveThe accuracy and precision of the new IRMA (Immediate Response Mobile Analysis System, Diametrics, Inc., St. Paul, MN) handheld blood gas analyzer was compared with that of two benchtop blood gas analyzers. The IRMA consists of a notebook-sized machine and disposable cartridges, each containing a pH, a CO2 and an O2 electrode, and provides bedside (point-of-care) blood gas analysis.MethodsA total of 172 samples (arterial and mixed venous) were obtained from 25 informed, consenting patients undergoing cardiopulmonary bypass. The pH, PCO2 and PO2 of each sample was determined on four blood gas analyzers: NOVA Statlabs Profile 5 (NOVA Biomedical, Waltham, MA), the ABL-50 (Radiometer, West Lake, OH), and two IRMA machines. Linear regression and bias +/- precision were determined, comparing each of the analyzers with the NOVA.ResultsAll three machines showed a similar, high degree of correlation with the NOVA for pH, PCO2, and PO2. The bias and precision of the IRMA machines compared with the NOVA was similar to that of the ABL compared with the NOVA for pH (NOVA:ABL -0.005 +/- 0.011; NOVA:IRMA 1 = 0.0026 +/- 0.025; NOVA:IRMA 2 = 0.0021 +/- 0.025), for PCO2 (NOVA:ABL = -1.4 +/- 1.3 mmHg; NOVA: IRMA 1 = -1.3 +/- 1.9 mmHg; NOVA: IRMA 2 = -1.2 +/- 2.1 mmHg) and PO2 (NOVA:ABL = 3.6 +/- 21.1 mmHg; NOVA:IRMA 1 = 3.4 +/- 19.9 mmHg; NOVA:IRMA 2 = 6.3 +/- 20.9 mmHg). The bias found for pH, PCO2, and PO2 was not affected by extremes of temperature (range 25.5-40 degrees C) or hematocrit (range 11-44%) for any machine.ConclusionsThe new technology incorporated in the IRMA blood gas analyzer provides results with an accuracy that is similar to that of benchtop analyzers, but with all of the advantages of point-of-care analysis.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…