    Effects of flow pattern, device and formulation on particle size distribution of nebulized aerosol.

    • Junhua Hu, Rui Zhang, Huimin Beng, Liangjun Deng, Qingjin Ke, and Wen Tan.
    • Institute of Biomedical and Pharmaceutical Sciences, Guangdong University of Technology, Guangzhou, Guangdong, China.
    • Int J Pharm. 2019 Apr 5; 560: 35-46.

    AbstractThe evaluation of particle size recommended in the pharmacopeias requires a constant flow rate, and the method for pediatric inhaled drugs is the same as for adult drugs. In this study, the aerosol concentration and particle size distribution (PSD) were measured under a realistic breathing pattern and constant flow. Two types of nebulizer (i.e., breath-enhanced nebulizer and vibrating-mesh nebulizer) and two formulations (i.e., budesonide suspension and albuterol solution) were chosen for comparison. The aerosol concentration under the realistic pattern was not constant, which was different from the result at constant flow. The changing trend of aerosol concentration varied with the operation process of each device. The aerosol concentration profile was similar between budesonide suspension and albuterol solution. As to the PSD, as inspiratory flow increased, the X50 (50% undersize) increased with all nebulizers but Omron microAir NE-U22 nebulizer. There was good agreement between X50 obtained under the realistic inhalation patterns and their equivalent average flow rates by Bland-Altman analysis, although the X50 obtained under the realistic inhalation pattern was greater than value at constant flow. The agreement of the two breath-enhanced jet nebulizers was better than that of the vibrating-mesh nebulizers. The X50 of budesonide was not equal to that of albuterol when using the same nebulizer. Interestingly, a significant difference was observed in the X50 and Span when comparing the results of PSD under adult and child breathing patterns. Furthermore, all vibrating-mesh nebulizers produced aerosol droplets having larger mean diameter and narrower size distribution than those of the air-jet nebulizers. We conclude that it will be more conducive to the evaluation of particle size to use a laser diffractometer under a realistic pattern and make up for the shortcomings of cascade impactors. The effects of flow pattern, nebulizer and formulation should be taken into account in the evaluation of the qualities of nebulizer products in pharmaceutical practice.Copyright © 2019 Elsevier B.V. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.