    The Impact of High-Flow Nasal Cannula Device, Nebulizer Type, and Placement on Trans-Nasal Aerosol Drug Delivery: An In Vitro Study.

    • Jie Li, Lucas Williams, and James B Fink.
    • Division of Respiratory Care, Department of Cardiopulmonary Sciences, Rush University, Chicago, Illinois. Jie_Li@rush.edu.
    • Respir Care. 2022 Jan 1; 67 (1): 181-8.

    BackgroundAerosol delivery via high-flow nasal cannula (HFNC) has been increasingly used in recent years. However, the effects of different HFNC devices, nebulizer types, and placement on aerosol deposition remain largely unknown.MethodsAn adult manikin with anatomically correct upper airway was used with a collection filter placed between the manikin's trachea and a breathing simulator, composed of a dual-chamber model lung driven by a critical care ventilator. Three HFNC device configurations were compared, with vibrating mesh nebulizer and small-volume nebulizer placed at the humidifier (inlet for Optiflow and outlet for Airvo 2) and proximal to the nasal cannula at gas flows of 10, 20, 40 and 60 L/min, in quiet and distressed breathing patterns. Albuterol (2.5 mg) was nebulized for each condition (no. = 3). The drug was eluted from the collection filter and assayed with ultraviolet spectrophotometry (276 nm).ResultsAt all settings, except when a nebulizer was placed proximal to the nasal cannula with the Optiflow and when the HFNC flow was set at 60 L/min, the vibrating mesh nebulizer generated a higher inhaled dose than did the small-volume nebulizer (all P < .05). With the exception of distressed breathing at an HFNC flow of 10 L/min, the inhaled dose with the vibrating mesh nebulizer placed at the humidifier was greater than with the vibrating mesh nebulizer placed proximal to the nasal cannula (all P < .05), Optiflow provided a higher inhaled dose than did Airvo 2 with either AirSpiral or 900PT501 circuits with the vibrating mesh nebulizer placed at the humidifier (all P < .05).ConclusionsDuring transnasal aerosol delivery, the vibrating mesh nebulizer generated a higher inhaled dose than did the small-volume nebulizer when the nebulizer was placed at the humidifier. With the vibrating mesh nebulizer placed at the humidifier and an HFNC flow > 10 L/min, the inhaled dose was higher than with the vibrating mesh nebulizer placed proximal to the nasal cannula, and the inhaled dose was higher with Optiflow than with Airvo 2.Copyright © 2022 by Daedalus Enterprises.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.