    Technical performance and reflection capacity of the anaesthetic conserving device--a bench study with isoflurane and sevoflurane.

    • Andreas Meiser, Martin Bellgardt, Javier Belda, Kerstin Röhm, Heinz Laubenthal, and Clemens Sirtl.
    • Klinik für Anaesthesiologie, St. Josef-Hospital, Ruhr-Universität Bochum, Gudrunstr. 56, 44791, Bochum, Germany. andreas.meiser@rub.de
    • J Clin Monit Comput. 2009 Feb 1;23(1):11-9.

    ObjectiveThe anaesthetic conserving device (AnaConDa), Sedana Medical, Sundbyberg, Sweden) facilitates administration of isoflurane or sevoflurane by liquid infusion. An anaesthetic reflector inside the device conserves exhaled anaesthetic and re-supplies it during inspiration. In this bench study, we examined the influence of infusion rates and ventilatory settings on the resulting anaesthetic concentrations on patient (C(pat)) and ventilator side of the reflector (C(loss)) to describe its technical performance.MethodsA Puritan Bennett 840 ICU ventilator (Pleasanton, US), AnaConDa, and a test lung (3 l-chloroprene-bag) were assembled. Infusion rates (IR, 0.2-50 ml h(-1)), respiratory rates (RR, 5-40 breaths min(-1)), and tidal volumes (V(T), 0.3, 0.5, and 1.0 l) were varied. C(pat) was measured via a thin catheter in the middle of the 3 l-bag in steady state (online data storage and averaging over >10 min). C(loss) was calculated from IR (to yield the volume of vapour per unit of time), and expired minute volume (in which the vapour is diluted) on the assumption that, in the steady state, input by liquid infusion equals output through the reflector.ResultsAt lower concentrations (C(pat) < 1 vol%) the ratio C(loss)/C(pat) was constant (R(C) = 0.096 +/- 0.012) for all combinations of IR, RR and V(T), both for isoflurane and sevoflurane. The device could efficiently reflect up to 10 ml vapour per breath (e.g. 2 vol% in 0.5 l). When exceeding this capacity, surplus vapour "spilled over" and R(C) markedly increased indicating decreased performance.ConclusionsThe triple product minute volume times R(C) times C(pat) describes anaesthetic losses through the reflector. It can easily be calculated as long as the 10 ml reflection capacity is not exceeded and thus R(C) is constant. Increased minute ventilation necessitates increasing the IR to keep C(pat) constant. When using large V(T) and high C(pat) "spill over" occurs. This effect offers some protection against an inadvertent overdose.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.