• J. Am. Coll. Cardiol. · Oct 2012

    Development of a novel echocardiography ramp test for speed optimization and diagnosis of device thrombosis in continuous-flow left ventricular assist devices: the Columbia ramp study.

    • Nir Uriel, Kerry A Morrison, Arthur R Garan, Tomoko S Kato, Melana Yuzefpolskaya, Farhana Latif, Susan W Restaino, Donna M Mancini, Margaret Flannery, Hiroo Takayama, Ranjit John, Paolo C Colombo, Yoshifumi Naka, and Ulrich P Jorde.
    • Division of Cardiology, Department of Medicine, Columbia University Medical Center, New York, New York 10032, USA. urielnir@gmail.com
    • J. Am. Coll. Cardiol. 2012 Oct 30;60(18):1764-75.

    ObjectivesThis study sought to develop a novel approach to optimizing continuous-flow left ventricular assist device (CF-LVAD) function and diagnosing device malfunctions.BackgroundIn CF-LVAD patients, the dynamic interaction of device speed, left and right ventricular decompression, and valve function can be assessed during an echocardiography-monitored speed ramp test.MethodsWe devised a unique ramp test protocol to be routinely used at the time of discharge for speed optimization and/or if device malfunction was suspected. The patient's left ventricular end-diastolic dimension, frequency of aortic valve opening, valvular insufficiency, blood pressure, and CF-LVAD parameters were recorded in increments of 400 rpm from 8,000 rpm to 12,000 rpm. The results of the speed designations were plotted, and linear function slopes for left ventricular end-diastolic dimension, pulsatility index, and power were calculated.ResultsFifty-two ramp tests for 39 patients were prospectively collected and analyzed. Twenty-eight ramp tests were performed for speed optimization, and speed was changed in 17 (61%) with a mean absolute value adjustment of 424 ± 211 rpm. Seventeen patients had ramp tests performed for suspected device thrombosis, and 10 tests were suspicious for device thrombosis; these patients were then treated with intensified anticoagulation and/or device exchange/emergent transplantation. Device thrombosis was confirmed in 8 of 10 cases at the time of emergent device exchange or transplantation. All patients with device thrombosis, but none of the remaining patients had a left ventricular end-diastolic dimension slope >-0.16.ConclusionsRamp tests facilitate optimal speed changes and device malfunction detection and may be used to monitor the effects of therapeutic interventions and need for surgical intervention in CF-LVAD patients.Copyright © 2012 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.