Infrared CO2 analyzer error: an effect of background gas (N2 and O2).
- R Arieli, O Ertracht, and Y Daskalovic.
- Israel Naval Medical Institute, Israel Defense Forces Medical Corps, Haifa 31080, Israel.
- J. Appl. Physiol. 1999 Feb 1;86(2):647-50.
AbstractThree infrared CO2 analyzers were tested for the effect of background gases: the Ametek CD-3A (Ametek, Thermox Instruments Division, Pittsburgh, PA), the Dräger Multiwarn P CO2 (Dräger, L ubeck, Germany), and the Servomex 1440 (Servomex, Crowborough, East Sussex, UK). Various CO2 concentrations were prepared with Wösthoff precision pumps (H. Wösthoff, Bochum, Germany). Calibration with a different background gas (O2 or N2) caused a similar but systematic error in the CO2 readings of all three analyzers. When the CO2 analyzers were calibrated with N2 as the background gas, the CO2 reading in an O2-enriched atmosphere was 8% lower than the true value. Conversely, calibration with O2 as the background gas resulted in a 10% overestimation of CO2 levels when N2 was the background gas. This error may be important in a few fields of respiratory physiology.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.