• J Hand Surg Am · May 2014

    Distal radius volar locking plate design and associated vulnerability of the flexor pollicis longus.

    • Roongsak Limthongthang, Abdo Bachoura, Sidney M Jacoby, and A Lee Osterman.
    • Philadelphia Hand Center, Department of Orthopaedic Surgery, Thomas Jefferson University, Philadelphia, PA; Department of Orthopaedic Surgery, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand. Electronic address: roongsak.lit@mahidol.ac.th.
    • J Hand Surg Am. 2014 May 1;39(5):852-60.

    PurposeFlexor pollicis longus (FPL) tendon rupture is a well-documented complication related to the use of distal radius volar locking plates (VLPs). The final common pathway of flexor tendon rupture appears to involve implants prominent at the watershed line. We hypothesized that significant differences in VLP prominence exist between various plate designs.MethodsTen fresh frozen specimens were dissected to identify the path of the FPL in relationship to the distal radius at the watershed line. Five VLP designs were fixed to each specimen based on their anatomic fit, and slid distally until the distal edge of the plate reached the watershed line. The position of each fixed plate was evaluated by fluoroscopy. We used a 3-dimensional laser scanner to create computer models. The total surface area of plate prominence volar to the watershed line and the prominent area beneath the FPL were measured in the axial plane using computer software.ResultsAt the watershed line, the FPL was located at 54% of the maximal width of the radius, as measured from its volar-ulnar corner. There were no significant differences in the location of plate fixation on lateral view radiographs according to the classification of Soong et al. The mean total surface area of plate prominence was 36 mm(2). The mean prominent area beneath the FPL was 10 mm2. Significant differences in plate prominence were noted for various designs.ConclusionsDespite optimal plate placement, various VLP designs were observed to have prominent profiles volar to the watershed line, to varying extents.Clinical RelevanceThe results raise concerns regarding interference between all of the analyzed VLP designs and the FPL. This study may help guide both implant design considerations and assist the surgeon in better understanding implant morphology as it relates to iatrogenic flexor tendon injury.Copyright © 2014 American Society for Surgery of the Hand. Published by Elsevier Inc. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.