• BMC anesthesiology · Oct 2021

    Comparative Study Observational Study

    Comparison of the resonance sonorheometry based Quantra® system with rotational thromboelastometry ROTEM® sigma in cardiac surgery - a prospective observational study.

    • Werner Baulig, Samira Akbas, Philipp K Schütt, Wolfgang Keul, Marija Jovic, Pascal Berdat, Stefanie von Felten, Klaus Steigmiller, Michael Thomas Ganter, and Oliver M Theusinger.
    • Department of Anesthesiology and Intensive Care Medicine, Klinik Im Park, Seestrasse 220, CH-8027, Zurich, Switzerland. werner.baulig@hirslanden.ch.
    • BMC Anesthesiol. 2021 Oct 28; 21 (1): 260.

    BackgroundMeasures of the sonorheometry based Quantra® viscoelastic hemostatic analyzer (HemoSonics, LCC, Charlottesville, VA, USA) were compared with corresponding results of the ROTEM® sigma device (Instrumentation Laboratory, Bedford, MA, USA).MethodsIn thirty-eight patients scheduled for elective cardiac surgery between December 2018 and October 2019, blood samples were taken after induction of anesthesia (sample 1) and after heparin neutralization (sample 2) and measured on Quantra (QPlus® Cartridge) and ROTEM sigma (ROTEM® sigma complete + hep Cartridge). Clot times and clot stiffness values were recorded. Clot stiffness values of ROTEM amplitudes (A in mm) were converted to shear modulus (G) in hectoPascal (hPa): G (hPa) = (5 x A)/(100-A). Additionally, time-to-results was recorded. Spearman rank test correlation and Bland Altman analysis were performed.ResultsClot stiffness parameters of the Quantra correlated strongly with corresponding measurements of the ROTEM with r = 0.93 and 0.94 for EXTEM A10 vs CS and r = 0.94 and 0.96 for FIBTEM A10 vs FCS for sample 1 and 2, respectively. Quantra clot time correlated strongly with ROTEM INTEM CT with r = 0.71 for sample 1 and r = 0.75 for sample 2. However, Bland Altman analysis showed no agreement in all compared assays of both methods. The median time to delivery of first and complete results was significantly shorter for Quantra (412 and 658 s) compared to ROTEM sigma (839 and 1290 s).ConclusionsThe Quantra showed a strong correlation with the ROTEM sigma for determining clot times and clot stiffness and the parameters assess similar aspects of clot development. However, these parameters are not directly interchangeable and implicate that separate cut-off values need to be established for users of the Quantra device. Word count: 278.Trial RegistrationThe study was retrospectively registered with ClinicalTrials.gov (ID: NCT04210830 ) at December 20th 2019.© 2021. The Author(s).

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…