J Diabetes Sci Technol · Nov 2020
Continuous Glucose Monitoring in Critically Ill Patients With COVID-19: Results of an Emergent Pilot Study.
- Archana R Sadhu, Ivan Alexander Serrano, Jiaqiong Xu, Tariq Nisar, Jessica Lucier, Anjani R Pandya, and Bhargavi Patham.
- Division of Endocrinology, Diabetes and Metabolism, Houston Methodist, Weill Cornell Medical College, Texas A&M Health Sciences Center, Houston, TX, USA.
- J Diabetes Sci Technol. 2020 Nov 1; 14 (6): 1065-1073.
BackgroundAmidst the coronavirus disease 2019 (COVID-19) pandemic, continuous glucose monitoring (CGM) has emerged as an alternative for inpatient point-of-care blood glucose (POC-BG) monitoring. We performed a feasibility pilot study using CGM in critically ill patients with COVID-19 in the intensive care unit (ICU).MethodsSingle-center, retrospective study of glucose monitoring in critically ill patients with COVID-19 on insulin therapy using Medtronic Guardian Connect and Dexcom G6 CGM systems. Primary outcomes were feasibility and accuracy for trending POC-BG. Secondary outcomes included reliability and nurse acceptance. Sensor glucose (SG) was used for trends between POC-BG with nursing guidance to reduce POC-BG frequency from one to two hours to four hours when the SG was in the target range. Mean absolute relative difference (MARD), Clarke error grids analysis (EGA), and Bland-Altman (B&A) plots were calculated for accuracy of paired SG and POC-BG measurements.ResultsCGM devices were placed on 11 patients: Medtronic (n = 6) and Dexcom G6 (n = 5). Both systems were feasible and reliable with good nurse acceptance. To determine accuracy, 437 paired SG and POC-BG readings were analyzed. For Medtronic, the MARD was 13.1% with 100% of readings in zones A and B on Clarke EGA. For Dexcom, MARD was 11.1% with 98% of readings in zones A and B. B&A plots had a mean bias of -17.76 mg/dL (Medtronic) and -1.94 mg/dL (Dexcom), with wide 95% limits of agreement.ConclusionsDuring the COVID-19 pandemic, CGM is feasible in critically ill patients and has acceptable accuracy to identify trends and guide intermittent blood glucose monitoring with insulin therapy.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.