    Multicenter reproducibility of quantitative susceptibility mapping in a gadolinium phantom using MEDI+0 automatic zero referencing.

    • Kofi Deh, Keigo Kawaji, Marjolein Bulk, Louise Van Der Weerd, Emelie Lind, Pascal Spincemaille, Kelly McCabe Gillen, Johan Van Auderkerke, Yi Wang, and Thanh D Nguyen.
    • Department of Radiology, Weill Medical College of Cornell University, New York, New York.
    • Magn Reson Med. 2019 Feb 1; 81 (2): 1229-1236.

    PurposeTo determine the reproducibility of quantitative susceptibility mapping at multiple sites on clinical and preclinical scanners (1.5 T, 3 T, 7 T, and 9.4 T) from different vendors (Siemens, GE, Philips, and Bruker) for standardization of multicenter studies.MethodsSeven phantoms distributed from the core site, each containing 5 compartments with gadolinium solutions with fixed concentrations between 0.625 mM and 10 mM. Multi-echo gradient echo scans were performed at 1.5 T, 3 T, 7 T, and 9.4 T on 12 clinical and 3 preclinical scanners. DICOM images from the scans were processed into quantitative susceptibility maps using the Laplacian boundary value (LBV) and MEDI+0 automatic uniform reference algorithm. Region of interest (ROI) analyses were performed by a physicist to determine agreement between results from all sites. Measurement reproducibility was assessed using regression, Bland-Altman plots, and the intra-class correlation coefficient (ICC).ResultsQuantitative susceptibility mapping (QSM) from all scanners had similar, artifact-free visual appearance. Regression analysis showed a linear relationship between gadolinium concentrations and average QSM measurements for all phantoms (y = 350x - 0.0346, r2 >0.99). The SD of measurements increased almost linearly from 32 ppb to 230 ppb as the measured susceptibility increased from 0.26 ppm to 3.56 ppm. A Bland-Altman plot showed the bias, upper, and lower limits of agreement for all comparisons were -10, -210, and 200 ppb, respectively. The ICC was 0.991 with a 95% CI (0.973, 0.99).ConclusionsQSM shows excellent multicenter reproducibility for a large range of susceptibility values encountered in cranial and extra-cranial applications on a diverse set of scanner platforms.© 2018 International Society for Magnetic Resonance in Medicine.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.