• Chest · Oct 2021

    Case Reports

    Oral Positive Expiratory Pressure Device for Excessive Dynamic Airway Collapse Caused by Emphysema.

    • Muhammad Ahsan Zafar, Ruchira Sengupta, Alister Bates, Jason C Woods, Christopher Radchenko, Francis X McCormack, and Ralph J Panos.
    • Department of Internal Medicine, Division of Pulmonary and Critical Care Medicine, University of Cincinnati College of Medicine, Cincinnati, OH. Electronic address: zafarmd@ucmail.uc.edu.
    • Chest. 2021 Oct 1; 160 (4): e333-e337.

    AbstractExcessive dynamic airway collapse (EDAC) contributes to breathlessness and reduced quality of life in individuals with emphysema. We tested a novel, portable, oral positive expiratory pressure (o-PEP) device in a patient with emphysema and EDAC. MRI revealed expiratory tracheal narrowing to 80 mm2 that increased to 170 mm2 with the o-PEP device. After 2-weeks use of the o-PEP device for 33% to 66% of activities, breathlessness, quality of life, and exertional dyspnea improved compared with minimal clinically important differences (MCID): University of California-San Diego Shortness of Breath questionnaire score declined 69 to 42 (MCID, ≥5), St. George's Respiratory Questionnaire score decreased 71 to 27 (MCID, ≥4), and before and after the 6-minute walk test Borg score difference improved from Δ3 to Δ2 (MCID, ≥1). During the 6-minute walk test on room air without the use of the o-PEP device, oxyhemoglobin saturation declined 91% to 83%; whereas, with the o-PEP device, the nadir was 90%. Use of the o-PEP device reduced expiratory central airway collapse and improved dyspnea, quality of life, and exertional desaturation in a patient with EDAC and emphysema.Copyright © 2021 American College of Chest Physicians. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…