• The Journal of pediatrics · Oct 1994

    A practical and reliable method of measuring blood pressure in the neonate by pulse oximetry.

    • M Langbaum and F G Eyal.
    • Department of Neonatology, Johns Hopkins Bayview Medical Center, Johns Hopkins School of Medicine, Baltimore, Maryland.
    • J. Pediatr. 1994 Oct 1;125(4):591-5.

    ObjectiveWe investigated the reliability of the plethysmographic waveform of the pulse oximeter to measure systolic blood pressure in sick neonates.MethodsFifty infants admitted to the neonatal intensive care unit, with indwelling arterial catheters placed for their ongoing care, were enrolled. Median gestational age was 31 weeks (range, 24 to 40 weeks), and the mean birth weight was 1711 gm (range, 546 to 3856 gm). Blood pressure was recorded by an oscillometric method as well as from a transducer connected to an arterial catheter. Additionally, pulse oximeter blood pressures were obtained by gradually inflating an appropriately sized blood pressure cuff in increments of 2 to 5 mm Hg, on the same extremity as the oximeter probe, until the waveform just disappeared. The cuff was then rapidly inflated another 20 mm Hg and then gradually deflated in increments of 2 to 5 mm Hg until the waveform reappeared on the oximeter screen display. The pulse oximeter blood pressures were calculated both as the blood pressure noted at disappearance of the pulse oximeter waveform and as the blood pressure noted by the average pulse oximeter blood pressure at the disappearance and reappearance of the waveform. The mean intraarterial systolic blood pressure was 54 mm Hg (range, 36 to 82 mm Hg).ResultsBlood pressures obtained by pulse oximetry showed a significantly better correlation with intraarterial measurements in comparison with those obtained by oscillometric instruments. Additionally, the limits of agreement (mean difference +/- 2 SD) between blood pressures obtained by intraarterial measurements and those obtained by pulse oximetry were within a clinically acceptable range as opposed to those obtained by the comparison of intraarterial and oscillometric methods.ConclusionMeasurements of blood pressure in the neonate by means of pulse oximetry waveform analysis are easily obtainable and more accurate than those obtained by the oscillometric method.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…