• J. Thorac. Cardiovasc. Surg. · Aug 1996

    Comparative Study

    Effects of failure of the right side of the heart and increased pulmonary resistance on mechanical circulatory support with use of the miniaturized HIA-VAD displacement pump system.

    • F Waldenberger, Y I Kim, S Laycock, B Meyns, and W Flameng.
    • Department of Cardiac Surgery, Katholieke Universiteit Leuven, Belgium.
    • J. Thorac. Cardiovasc. Surg. 1996 Aug 1;112(2):484-93.

    AbstractThis experimental study was designed to assess the influence of failure of the right side of the heart or pulmonary hypertension, or both, on the performance of a novel miniaturized left ventricular assist device. In small-sized dogs (n = 50) ischemic global left ventricular failure was induced and support was provided by the HIA-VAD displacement pump (stroke volume 10 or 25 ml) installed as a left ventricular assist device. In three groups of animals (n = 10 each) pulmonary hypertension was created before induction of global left ventricular failure. During left ventricular assist device support temporary ischemic failure of the right side of the heart was induced in four groups of animals (n = 10 each). In the group subjected to left ventricular failure, support with the left ventricular assist device, and right ventricular failure during left ventricular assist, left atrial pressure and cardiac index were significantly lower than in the group subjected to left ventricular failure and left ventricular assist alone (2 +/- 6 versus 11 +/- 6 mm Hg and 1.6 +/- 0.4 versus 1.0 +/- 0.4 L/(min/m2), respectively, p < 0.05). In the group subjected to pulmonary hypertension, left ventricular failure, and left ventricular support, left atrial pressure dropped to values near zero but cardiac index remained unaltered as compared with results with the same regimen without pulmonary hypertension. However, when right ventricular failure was added (that is, pulmonary hypertension, left ventricular failure, left ventricular support, and right ventricular failure during support with the left ventricular assist device) left atrial pressure dropped to negative values (p < 0.05) and cardiac index progressively deteriorated. When, in an additional group of dogs, biventricular support was installed in the latter regimen, circulation was initially well supported but oxygenation deteriorated in 60% of cases. We conclude that (1) adequate right ventricular function was indispensable during support with the left ventricular assist device, (2) the combination of pulmonary hypertension and right ventricular failure led to the "low left ventricular assist device output syndrome," and (3) biventricular mechanical support in the presence of pulmonary hypertension may be complicated by the alveolar leakage syndrome.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.