• J Clin Monit Comput · Oct 2022

    Lag times to steady state drug delivery by continuous intravenous infusion: direct comparison of peristaltic and syringe pump performance identifies contributions from infusion system dead volume and pump startup characteristics.

    • Lauren E Gibson, Anders S Knudsen, David Arney, Hao Deng, Nathaniel M Sims, and Robert A Peterfreund.
    • Department of Anesthesia, Critical Care, and Pain Medicine, Massachusetts General Hospital, 55 Fruit Street, Boston, MA, USA. legibson@partners.org.
    • J Clin Monit Comput. 2022 Oct 1; 36 (5): 1489-1498.

    AbstractTime lags between the initiation of a continuous drug infusion and achievement of a steady state delivery rate present an important safety concern. At least 3 factors contribute to these time lags: (1) dead volume size, (2) the ratio between total system flow and dead volume, and (3) startup delay. While clinicians employ both peristaltic pumps and syringe pumps to propel infusions, there has been no head-to-head comparison of drug delivery between commercially available infusion pumps with these distinct propulsion mechanisms. We quantified the delivery of a model drug by peristaltic and syringe pumps at clinically relevant flow rates using spectrophotometric absorbance. Delivery curves were modeled and compared, and the time required to reach 5% (T5), 50% (T50), and 95% (T95) of the intended delivery rate was reported. The ability to overcome the combined effects of startup delay and dead volume differed between syringe and peristaltic pumps. T5, T50, and T95 were shorter for the peristaltic pump at higher flow rates. T50 and T95 were shorter for the syringe pump at lower flow rates. The ability to overcome the effects of dead volume was overall similar between the syringe and peristaltic pumps, as was the response to consecutive changes in drug infusion rates. Startup delay and dead volume in carrier-based infusion systems cause substantial time lags to reaching intended delivery rates. Peristaltic and syringe pumps are similarly susceptible to dead volume effects. Startup performance differed between peristaltic and syringe pumps; their relative performance may be dependent on flow rate.© 2022. The Author(s), under exclusive licence to Springer Nature B.V.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…