Biomechanical Evaluation of Extramedullary Versus Intramedullary Reduction in Unstable Femoral Trochanteric Fractures.
- Tadashi Kawamura, Hiroaki Minehara, Ryo Tazawa, Terumasa Matsuura, Rina Sakai, and Masashi Takaso.
- Department of Orthopaedic Surgery, Kitasato University School of Medicine, Minami-ku, Sagamihara, Kanagawa, Japan.
- Geriatr Orthop Surg Rehabil. 2021 Jan 1; 12: 2151459321998611.
IntroductionThe failure rate of operations involving the cephalomedullary nail technique for unstable femoral trochanteric fractures is 3-12%. Changing the reduction strategy may improve the stability. This study aimed to confirm whether reducing the proximal fragment with the medial calcar contact, as opposed to utilizing an intramedullary reduction, would improve the stability of such fractures.Materials And MethodsThe unstable femoral trochanteric fracture model was created with fixation by cephalomedullary nails in 22 imitation bones. The 2 reduction patterns were as follows: one was with the proximal head-neck fragment external to the distal bone in the frontal plane and anterior in the sagittal plane as "Extramedullary," while the other was the opposite reduction position, that is, bone in the frontal plane and sagittal plane as "Intramedullary." We evaluated the tip-apex distance, compression stiffness, change in femoral neck-shaft angle, amount of blade telescoping, and diameter of the distal screw hole after the compression test. Statistical analysis was conducted using the Mann-Whitney U test.ResultsNo significant differences were seen in compression stiffness (p = 0.804) and femoral neck-shaft angle change (p = 0.644). Although the "Extramedullary" tip-apex distance was larger than the "Intramedullary" distance (p = 0.001), it indicated clinically acceptable lengths. The amount of blade telescoping and the distal screw hole diameter were significantly larger in "Intramedullary" than in "Extramedullary" (p < 0.001, p = 0.019, respectively). Our results showed that "Intramedullary" had significantly larger blade telescoping and distal screw hole diameters than "Extramedullary," and contrary to our hypothesis, no significant differences were seen in compression stiffness and femoral neck-shaft angle change.ConclusionsAs opposed to the "Intramedullary" reduction pattern, the biomechanical properties of the "Extramedullary" reduction pattern improved stability during testing and decreased sliding.© The Author(s) 2021.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.