    Pump in Parallel-Mechanical Assistance of Partial Cavopulmonary Circulation Using a Conventional Ventricular Assist Device.

    • Pranava Sinha, Nina Deutsch, Kanishka Ratnayaka, Dingchao He, Murfad Peer, Mustafa Kurkluoglu, Mark Nuszkowski, Erin Montague, Gerald Mikesell, David Zurakowski, and Richard Jonas.
    • ASAIO J. 2018 Mar 1; 64 (2): 238-244.

    AbstractMechanical assistance of systemic single ventricle is effective in pulling blood through a cavopulmonary circuit. In patients with superior cavopulmonary connection, this strategy can lead to arterial desaturation secondary to increased inferior caval flow. We hypothesized that overall augmentation in cardiac output with mechanical assistance compensates for the drop in oxygen saturation thereby maintaining tissue oxygen delivery (DO2). Bidirectional Glenn (BDG) was established in seven swine (25 kg) after a common atrium had been established by balloon septostomy. Mechanical circulatory assistance of the single ventricle was achieved using an axial flow pump with ventricular inflow and aortic outflow. Cardiac output, mean pulmonary artery pressure (PAP), common atrial pressure (left atrial pressure [LAP]), arterial oxygen saturation (SaO2), partial pressure of arterial oxygen (PaO2), and DO2 were compared between assisted and nonassisted circulation. Significant augmentation of cardiac output was achieved with mechanical assistance in BDG circulation (BDG: median [interquartile range {IQR}], 0.8 [0.9-1.15] L/min versus assisted BDG: median [IQR], 1.5 [1.15-1.7] L/min; p = 0.05). Although oxygen saturations and PaO2 trended to be lower with assistance (SaO2; BDG: median [IQR], 43% [32-57%]; assisted BDG: median [IQR], 32% [24-35%]; p = 0.07) (PaO2; BDG: median [IQR], 24 [20-30] mm Hg; assisted BDG: median [IQR], 20 [17-21] mm Hg; p = 0.08), DO2 was unchanged with mechanical assistance (BDG: median [IQR], 94 [35-99] ml/min; assisted BDG: median [IQR], 79 [63-85] ml/min; p = 0.81). No significant change in the LAP or PAP was observed. In the setting of superior cavopulmonary connection/single ventricle, the systemic ventricular assistance with a ventricular assist device (VAD) leads to increase in cardiac output. Arterial oxygen saturations however may be lower with mechanical assistance, without any change in DO2.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.