Randomized Controlled Trial Comparative Study
Hemodynamic comparison between Trifecta and freestyle aortic valve during exercise in patients with small aortic root.
- Giordano Tasca, Paola Redaelli, Beatrice Riva, Caterina Chiara De Carlini, Elisabetta Lobiati, and Amando Gamba.
- Department of Cardiovascular, Operative Unit of Cardiac Surgery, Ospedale "A. Manzoni" di Lecco, Lecco, Italy; Department of Electronic, Information and Bioengeering, Politecnico di Milano, Milan, Italy.
- J Card Surg. 2015 May 1; 30 (5): 400-4.
BackgroundPatients with a small aortic annulus, that is ≤ 23 mm, constitute a challenge for the surgeon, because they are at high risk of patient-prosthesis mismatch. Stentless valves provide better hemodynamic performance at rest and during exercise than stented valves, and are advocated in this group of patients. A new-generation stented valve, the Trifecta (St. Jude), has recently become available with improved hemodynamics. The aim of this study was to compare the hemodynamic performance of Freestyle (Medtronic) and Trifecta at rest and during exercise in patients with a small aortic annulus.MethodsFrom September 2012 to September 2014, 22 patients with a native aortic annulus ≤ 23 mm underwent ergometric stress testing one year after aortic valve replacement with either a Trifecta (12 patients) or a Freestyle (10 patients) bioprosthesis as part of a randomized study.ResultsThe mean gradient at rest was 6.0 ± 2.3 mmHg for Trifecta and 4.3 ± 3.5 for Freestyle (p = 0.213). The mean gradient at peak of exercise was 9.7 ± 3.4 mmHg for Trifecta and 7.4 ± 5 mmHg for Freestyle (p = 0.243). No significant differences were found between the two prostheses regarding other hemodynamic parameters: effective orifice area, velocity index, and performance indexes.ConclusionBoth the stented Trifecta and stentless Freestyle prostheses provide excellent hemodynamic results during physical stress in patients with a small aortic annulus. Our study confirms that Trifecta implantation results in low gradients at rest and during exercise and that the performance of Trifecta is similar to that of a stentless valve.© 2015 Wiley Periodicals, Inc.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.